MI 2883 Energy Master
Operating the instrument
90
DURATION
Duration of RMS record, which will be captured after
treshold value is reached.
THRESHOLD
Threshold value expressed in % of nominal voltage,
which will trigger recording of signalling event.
Table 3.81: Keys in Signalling setup screen
Enters or exits a submenu to set signalling frequency.
Toggles between given parameters.
Changes selected parameter.
Returns
to the “MEASUREMENT SETUP” submenu.
3.19.5
Rapid voltage changes (RVC) setup
RVC is a quick transition in RMS voltage occurring between two steady-state
conditions, and during which the RMS voltage does not exceed the dip/swell thresholds.
An voltage is in a steady-state condition if all the immediately preceding 100/120
U
Rms
(½)
values remain within an set RVC threshold from the arithmetic mean of those 100/120
U
Rms
(½)
(100 values for 50 Hz nominal and 120 values for 60 Hz). The RVC threshold is
set by the user according to the application, as a percentage of
U
Nom
, within 1 ÷ 6 %.
See section 5.1.14 for details regarding RVC measurement. See section 3.14 for
instructions how to start recording.
Figure 3.62: RVC setup screen
Table 3.82: Description of RVC setup
Nominal voltage
Indication of type (L-N or L-L) and value of nominal
voltage.
RVC THRESHOLD
RVC threshold value expressed in % of nominal voltage for
steady state voltage detection.
RVC HYSTERESIS
RVC hysteresis value expressed in % of RVC threshold.
Table 3.83: Keys in RVC setup screen
Toggles between given parameters.
ENTER