MI 2142 AlphaPAT Measurements
29
6.6.1 How to perform the measurement
Select
Polarity test
in
Single test menu
by using and cursor keys and press
ENTER key to confirm.
Polarity test menu
is displayed.
Polarity test menu
Press START key to perform test.
After performing the test PASS or FAIL indication appears on the display. In case of
FAIL result the detected fault is displayed.
Examples of polarity test results
Press START key to repeat the test or press ESC key to return to the
Single test
menu
.
Note:
Consider any displayed warning before starting test!
6.7 True RMS current measurement
This function enables measuring of AC currents in a wide range from 1mA up to 25A.
Typical applications are:
measuring leakage currents through PE conductor in permanently installed
appliances,
measuring load currents through L or N conductor in permanently installed
appliances,
measuring leakage currents through parallel current paths, etc.
6.7.1 How to perform TRMS clamp current measurement
In order to measure TRMS current by using current clamp connect the clamp to the
instrument CLAMP connector first.