background image

A 1588   

                               

            

 Multi range AC current clamps  

 

 

  Deterioration of insulation (because of pollution, moisture, corrosion). This is causing 

a gradual increase of the resistive leakage current. 

  Faults in electronic equipment. 

  If the neutral and PE wiring are connected together anywhere in the installation this 

can result in an improper current distribution through the neutral and PE conductors. 

Note 

Metrel Application note: 

Measuring of leakage current

 contain many practical aspects and 

examples of leakage current measurement.  

Maintenance 

Inspection 

To maintain operator safety and ensure reliability of the current clamp

 

it is good practice to 

inspect it on a regular basis. Check that the enclosure and optional connection are without 
defects such as scratches or breaks.  
Jaw surface must be clean. Pollution on jaw surfaces reduces the current clamp sensitivity. 

 

Cleaning 

Use a soft cloth moistened with soapy water or alcohol to clean non-metallic surface of the 
current clamps

 

and leave them to dry totally before using it.  

 

!

 

Do not use liquids based on petrol or hydrocarbons! 

!

 

Do not spill cleaning liquid over the current clamps! 

 

To clean jaw cut surfaces use slightly oiled soft cloth. 

 

Service And Calibration 

It is essential that your clamp is regularly calibrated in order to guarantee the technical 
specification listed in this User manual. We recommend 2-year calibration interval. Metrel 
encloses an original calibration certificate with every new instrument and clamp. 
For recalibration and repairs under or out of warranty time please contact your distributor for 
further information. 

 

!

 

Only a competent, authorized person is allowed to carry out service intervention!

 

Technical Specifications 

General  

Safety specification 

Over voltage category: 600 V CAT II,  
Pollution degree: 2 
Double insulation 
 

Environment conditions 

Working temperature: -10 

0

C ÷ 50 

0

Storage temperature: -20 

0

C ÷ 70 

0

Humidity: 0 % ÷ 85 %,  
Linearly decreasing for T > 35 

0

Altitude: 

 2000 m 

 

Applied standards 

Mechanical data 

Jaw opening: 40 mm 
 

Maximum conductor sizes: 

Cable: < 50 mm 
Bar:  1 bar 50 mm x 5 mm,  
 

4 bars 30 mm x 5 mm 

 
Housing flammability: UL94 

– V-1 

Dimension: 177 mm x 53 mm x 23 mm 
Weight: ~160 g 
 

Содержание A 1588

Страница 1: ...Multi range AC crrent clamps 0 5A 5A 50A A 1588 Instruction manual Version 1 0 Code no 20 752 829...

Страница 2: ...el metrel si 2018 METREL Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations No part of...

Страница 3: ...MENT 7 POWER MEASUREMENT 8 LEAKAGE CURRENT MEASUREMENT 8 Measurement Of Leakage Current By Direct Method 8 Measurement Of Differential Current And or Current Difference 8 Determining Problems In Light...

Страница 4: ...be touched during measurement Symbol on the current clamp indicates the possibility of a hazardous live condition if the operator ignores the required safety measures If the current clamps are used i...

Страница 5: ...885 Power Master MI 2892 Clamps have three current ranges 0 5A 5A and 50A which are selected directly by the connected instrument Embedded electronic module is powered directly from the connected inst...

Страница 6: ...Master MI 2892 Step 2 Switch on the measuring instrument Set Smart clamps and range on instrument s measuring setup For further information see Instruction manual of measuring instrument Step 3 Clamp...

Страница 7: ...rrent primary current Is Measured current current transformer secondary current N2 Number of secondary turns N2 1000 U I Output voltage as function of current U k Im Factor k depends on selected curre...

Страница 8: ...e active embraced conductors If no current is leaking to earth the sum of the currents passing through active conductors must be exactly zero regardless of the load currents If a leakage current is fl...

Страница 9: ...ve them to dry totally before using it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the current clamps To clean jaw cut surfaces use slightly oiled soft cloth S...

Страница 10: ...Frequency range 40 Hz 700 Hz 0 50 phase error 700 Hz 2500 Hz 30 phase error Continuity of measurements 100 A continuous 120 A 40 min 20 min intermitted Load impedance 1 k Working voltage 600 Vrms Inf...

Страница 11: ...and industrial use Immunity Intended for use in industrial locations but can also be used for domestic purposes Performance criteria B with addition to the tables bellow Table 1 Immunity to radiated...

Отзывы: