Flex current clamps A 1501, A 1502 and A 1503 

 

7

 

  It is very important that the conductor is in the centre and perpendicular to the 

current censor as much as possible, in order to minimise position 
measurement error.  

  Minimise influence of adjacent current-carrying conductors, and measure at 

the point where they are far away from each other. 

  Make sure that the arrow marked on the clamp coupling points toward the 

correct orientation for correct phase.  

  Keep the clamp coupling more than 2.5 cm (1 in) away from conductor.  

 

2. Connect Flexible Current Clamps to the desired input on the measuring instrument 

 

Figure 2: 3-phase 4-wire system 

3. Select Phase and Neutral clamps to 

Smart Clamp

 in Connection setup menu on 

your instrument. 

   

 

4. Select appropriate Current clamp range. 

5. 

Observe the current value and waveform on the instrument’s display. If desired, 
select the lower range on the measurement instrument. 

 

Содержание A 1501

Страница 1: ...Flex current clamps A 1501 A 1502 and A 1503 Instruction Manual Version 1 1 1 Code No 20 752 759...

Страница 2: ...5 49 226 E mail metrel metrel si http www metrel si Mark on your equipment certifies that it meets European Union requirements for EMC LVD ROHS regulations 2017 Metrel No part of this publication may...

Страница 3: ...tions 6 5 1 Power Quality Measurement 6 5 2 Energy and Power Measurement 6 6 Connecting and using clamps in the field 6 6 1 Effective current measuring range 8 7 Technical specifications 9 7 1 A 1501...

Страница 4: ...30 A 900 A 160 A 9000 A Embedded electronic module provides automatic clamp recognition for Metrel instruments and current range switching Module is powered directly from attached instrument so clamp...

Страница 5: ...ith voltages higher than 1000 V CAT III or 600 V CAT IV De energize the installation on which current will be measured or adopt safe operating procedures during application and removal of the current...

Страница 6: ...appliances machines etc High precision see Figure 5 and Figure 9 and wide measurement range can cover most of practical LV current measurements 5 2 Energy and Power Measurement Current clamps have sma...

Страница 7: ...that the arrow marked on the clamp coupling points toward the correct orientation for correct phase Keep the clamp coupling more than 2 5 cm 1 in away from conductor 2 Connect Flexible Current Clamps...

Страница 8: ...asured value reading and includes all relevant errors residual noise clamp accuracy position sensitivity and instrument accuracy Effective measuring range 3A 30A 300A 3000A 30000A 3000A range Error 5...

Страница 9: ...0 2 50 Hz 3 1 5 kHz External magnetic field rejection in reference to adjacent conductor Min 40 dB 1 of adjacent conductor current 4 Electrical Safety Double insulation EN 61010 CAT III 1000 V up to...

Страница 10: ...cy Accuracy 30Arms Figure 8 Phase error 30Arms 0 00 0 20 0 40 0 60 0 80 1 00 1 20 1 40 0 1 1 10 Accuracy of m v RMS Current A 0 00 0 02 0 04 0 06 0 08 0 10 0 12 0 10 1 00 10 00 Phase error degrees RMS...

Страница 11: ...al current noise 60A range 0 24 A 600A range 0 6 A 6000A range 6 A 3 Phase error 0 2 50 Hz 3 1 5 kHz Electrical Safety Double insulation EN 61010 CAT III 1000 V up to 3000 m Pollution degree 2 Working...

Страница 12: ...y 60A Figure 12 Phase error 60 A 0 00 0 10 0 20 0 30 0 40 0 50 0 60 0 70 0 50 100 150 Accuracy of m v RMS Current A 0 00 0 02 0 04 0 06 0 08 0 10 0 12 0 10 1 00 10 00 100 00 Phase error degrees RMS Cu...

Страница 13: ...g it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the device 9 Service For repairs under or out of warranty time please contact your Metrel distributor for furt...

Отзывы: