No load voltage readout: (a.c. peak or d.c. peak)
Range Resolution
Accuracy
0.0 V
199.9 V
0.1 V
(5 % of r 5 digits)
No load voltage readout: (a.c. r.m.s.)
Range Resolution
Accuracy
0.0 V
139.9 V
0.1 V
(5 % of r 5 digits)
Measurement circuit according to EN60974 – Figure 4.
Pass levels*………. d.c. 113 V peak / a.c. 68 V peak and 48 V r.m.s.
d.c. 113 V peak / a.c. 113 V peak and 80 V r.m.s.
d.c. 141 V peak / a.c. 141 V peak and 100 V r.m.s.
none.
Test terminals ....................... Test socket (A 1422: 16A-5p, 32A-5p or 16A-3p) and
terminals W1 / W2