A 1322 / A 1422
Measurements
21
5.5 Touch leakage current measurement
Refer to chapter
Measurements – Single tests for appliances
, paragraph
Touch leakage current
in master instrument user manual for detailed information.
Procedure:
Item Comment
Prepare test system.
See
chapter 4.3 Working
with the adapter.
Connect power supply plug of tested appliance into
proper test socket.
Connect probe to accessible isolated conductive
part of the appliance.
See
figure below.
Prepare and start measurement on the master
instrument.
Wait until measurement is finished or stop the
measurement.
Check result(s).
Store result(s).
Optional
Figure 5.5:
Touch leakage measurement
Warning:
During the test, DUT is powered as for normal operation. Pay attention on
possible electric shock, and danger of rotating/moving parts!
Note:
Consider any displayed warning before starting measurement!