OPERATION
Thermometer
icon
Figure 1 Thermometer icon on screen
The instrument beeps, when the sample temperature is too high, indicating there is a danger
of condensing water in the sample chamber and on the sensor. If this warning appears while
sampling, remove the PAWKIT, place the cup lid on the sample, and wait until the sample has
reached ambient temperature before attempting to read again.
If the sample is colder than the ambient temperature of the PAWKIT, the accuracy of the
reading after 5 min may be questionable. Wait until the sample temperature is similar to that
of the PAWKIT.
2.1.2 TAKING A MEASUREMENT
1. Open the PAWKIT by holding the case near the LCD with one hand and pulling down on
the plastic sensor cover tab with the other hand.
Snap cover
tab in place
Cover tab
Figure 2 Open the PAWKIT
The sensor cover rotates and snaps into the open position.