background image

Downloading Data

1. Visit 

zentracloud.com 

and create an account.

2.  Create an organization in the top left of the screen.

3.  Click “Add new” at the bottom left to create a new site.

4.  Click “Add new” again to create a plot within the site.

5.  Click “Add new” a third time and select “Device”.

6.  Click “Setup”, then “Create New”, then “Em50G subscription”.

7.  Enter the Device ID and the Password found on the bright orange  

card that came with the Em50G.  Both the ID and password are  

case sensitive.  Click “Subscribe”.

8.  The next time you would like to see new data, you can launch 

ZENTRA Cloud, click on the device in the device tree, and select the 

either “Map”, “List”, or “Details” to view your data. Download your 

data by pressing the “Download” icon on any of these pages.

]

Em50G Logger Configuration

1.  Take the brightly colored card containing your Device ID and put it in 

a safe place.  You will need this number and password to download 

your data from the METER Data Service.

2. If ECH2O Utility is not launched, launch ECH2O Utility.

3.  Connect the Em50G to your computer using the included USB cable.

4.  In the “Connect via” dropdown menu, select the COM port containing 

either “Decagon UCA” or “METER UCA”.

5.  Click the “Connect” button in ECH2O Utility.

7.  Pick your measurement interval from the dropdown menu, and 

choose the sensors plugged into each port.

Installing the Em50G in the field.

1.   Before installing your sensors and loggers, check the quality of 

the cellular communication between the Em50G and local cellular 

towers at the location you would like to monitor.

a.  Insert the included 5AA batteries into the Em50G and attach 

the Em50G antenna.

b. Open ECH2O Utility on your field computer and connect the 

Em50G to your computer using the included USB cable.

c.  In the “connect via” dropdown menu, select the COM port 

containing either “Decagon UCA” or “METER UCA”.

d.  Click the “Connect” button in ECH2O Utility.

f.   In ECH2O Utility, choose “Cellular Test”.

g.  Click the “Test” button.  The communication test takes          

about 30 s, so please be patient.  If cellular coverage is low, try 

moving (even just a few meters) to a new location and repeating 

the test until you have at least one star of communication.

2.  Install your sensors and attach the Em50G in an upright position 

to a firm post or PVC pipe, using the included zip ties or similar 

fasteners.

3.  Install the data logger in a location that will ensure that the solar 

panel receives the maximum amount of sunlight. In the northern 

hemisphere, this will typically mean that the data logger will face 

the south.  In the southern hemisphere, this will typically mean that 
the data logger will face the north.”

NOTE:  Em50/Em50G solar data loggers come with batteries preinstalled.  Remove 

battery pull-tab to activate batteries. Use only nickel-metal hydride (NiMH) batteries 

with the solar Em50/Em50G.

Отзывы: