VTI Instruments Corp.
152
5)
In the
Reference Instruments
section, select the appropriate instrument available on the
GPIB/LAN network for the calibrator, external DMM, and function generator. This
information can also be obtained from LAN/GPIB Gateway’s web page. Refer to the
LAN/GPIB Gateway’s user’s manual for more information.
F
IGURE
9-9:
D
EVICE
S
ELECTION AND
T
EMPERATURE
F
IELDS
6)
Once the instruments have been selected, click the
Connect
button. If the connection is
successful, the SFP will display the 5700A’s calibration information. Click
OK
to proceed.
Enter the DMM’s serial number when prompted.
7)
In the
Temperature
section, enter the ambient temperature, in °C, into the
External
field, as
shown in Figure 9-9.
8)
In the
Calibration Status
section, click the
Initiate
button.
9)
As the calibration process progresses, prompts will appear indicating the connections that
need to be made between the EX1200A DMM and the reference standards.
NOTE
The calibrator’s “Aux Current Out” must be connected to the DUT’s “Input I” jack at all times.
10)
Upon completion of calibration, a window will appear indicating whether any failures
occurred. Save the test logs for future reference.
11)
Copy the cal constants to a file for future reference.
P
ERFORMING
V
ERIFICATION
When verification is run, the user can generate a report that indicates the measured and expected
values for the points measured during the calibration process. Performing verification before and
after calibration can provide the user valuable information regarding the stability of an instrument.
Prior to performing verification, ensure the following:
The EX1200A DMM (device under calibration), the function generator, the DMM, and the
calibrator/amplifier must all be powered up for a minimum of one hour before calibration
The Agilent 3458A “Auto Cal All” procedure must be performed prior to verification.
The 5700A’s “Cal Zero” procedure must be performed prior to verification.
Ensure that the “I FUSE” is installed in the EX1200A DMM.
Verification Procedure
1)
Follow steps 1) through 3) in the Performing Calibration section.
2)
From the Calibration Functions section of the Calibration page, select Verification from the
drop down menu.
3)
Follow steps 5) through 9) in the Performing Calibration section.
4)
When the test is finished, a popup window will tell you if there were any failures. Save the
test logs and test results.
5)
Copy the test logs and test results. Use these values to make the calibration report.
Device Drop Menu
Enter Ambient Temperature
Содержание EX1200A SERIES
Страница 16: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM...
Страница 17: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM...
Страница 32: ......
Страница 50: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use...
Страница 63: ...www vtiinstruments com EX1200A Series Preparation for Use 63...
Страница 64: ......
Страница 72: ......
Страница 106: ...VTI Instruments Corp 106 EX1200A Triggering...
Страница 108: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL...
Страница 142: ...VTI Instruments Corp 142 EX1200A Web Page Operation...
Страница 144: ...VTI Instruments Corp 144 EX1200A Series Programming...
Страница 154: ......
Страница 160: ......