Basic Detector Operation
Octane Basic User Guide
Page 18 of 25
4.5
LIGHT ELEMENT OPERATION
1.
Check that the geometry is correct when collecting spectra. For example, an incorrect tilt used,
giving an incorrect Take Off Angle, will affect the Quant results. You can also correct this after
the data has been collected by editing the spectra file parameters.
2.
Check that you are using a well-defined peak when quantifying. Avoid using peaks that are
“buried” in the background.
3.
Use one of the longer amp times when specifically looking for light energy elements. They
typically provide better light element performance.
4.
Use low accelerating voltage when specifically looking for light energy elements.
3.
Make sure the samples are flat and homogeneous when possible.
4.
Collect for sufficient time to improve the statics when the peaks are very small, for example
when looking at Boron.
4.6
FREQUENTLY ASKED QUESTIONS (FAQ)
Q1. Should I leave the detector powered on all of the time?
A1. In most laboratory conditions (stable power, temperature controlled), leaving the detector powered
on all of the time is fine.
Q2. Should I leave the detector cold all of the time?
A2. The detector will reach operating temperature in about two minutes, so it does not need to be left
cooled when not in use. Spectra can be collected after the detector reaches its operating temperature,
but for optimal performance and stable peak positions, it may be best to wait about 60 minutes. It is
okay to leave the detector cold when the microscope chamber is left under vacuum. There is a user
preference option in the software to automatically turn the cooling off after inactivity, in case the user
prefers to turn the cooling off automatically.
Q3. The software displays a message that the detector temperature is not changing, or a
communication problem with the detector. What should I do?
A3. In this case, power off the detector using the switch on the small black power “brick” for the
detector then reboot the PC. Then power the detector back on after Windows has started.
Q4. How often should I calibrate?
A4. Many labs only calibrate their systems during a regularly scheduled service preventative
maintenance visit. The detector will need to be calibrated in some cases when then spectrum peaks do
not match the blue line modeled peaks. For this case, a calibration will bring the peaks into alignment
with the modeled line. In other cases, labs may want to set a regular calibration schedule any time from
weekly to monthly.
Q5. What count rate should I use to calibrate?
A5. We recommend using a copper and aluminum sample with a primary beam at ≥ 20 kV. Adjust the
sample position so both Copper and Aluminum can be seen on the same screen. Set the position so the
Copper and Aluminum Kα peaks are similar in height, but with the Al peak larger than the CuK peak by
20 to 50%. Set the beam conditions so the Dead Time is ≤ 40% as you should use during normal use.
Содержание Apollo XF Octane
Страница 1: ...Apollo XF XLT Octane Basic User Guide Rev 1 5 June 24 2016...
Страница 4: ...Contents Octane Basic User Guide Page 4 of 25 This page is intentionally left blank...
Страница 6: ...Introduction Octane Basic User Guide Page 6 of 25 This page is intentionally blank...
Страница 12: ...Installation Requirements Octane Basic User Guide Page 12 of 25 This page is intentionally blank...
Страница 23: ...Basic System Cabling Octane Basic User Guide Page 23 of 25 This page is intentionally blank...
Страница 24: ...Basic System Cabling Octane Basic User Guide Page 24 of 25 This page is intentionally blank...