3-30 | Model 920 Multi-Gas Analyzer
Purged Analyzers
The Electronics Enclosure and its Optical Bench and Measuring Cell
Extension are continuously purged with Instrument Air to adequately
dilute any flammable gases that may have been released into these areas.
This function effectively prevents an internal explosion when the ana-
lyzer is powered up and operating under normal sampling conditions.
AMETEK analyzers that require purged enclosures in hazardous locations
use an Expo Technologies MiniPurge
®
System With eTimer to perform
this function.
The purge air supply is normally obtained from the Instrument Air supply
connected directly to the Expo MiniPurge
®
System. The required purge
air supply pressure for your analyzer is indicated on Final “As-Built”
drawings in the analyzer Documentation Package.
After the Instrument Air supply has been connected to the MiniPurge
®
system, the
ALARM/PRESSURIZED
indicator on the MiniPurge
®
system
(Figure 3-9) turns from RED (Alarm condition) to GREEN (Pressurized).
Next, the system’s eTimer begins its purge timing cycle, and the four
PURGE TIMING
indicator LEDs will flash (YELLOW) sequentially, each
for 25 % of the total purge time (minimum 15 minutes). During this stage,
the MiniPurge
®
system will purge any flammable gases that may have
entered the Electronics Enclosure while its door was open, and then pres-
surize the enclosure. After the timing sequence is completed (LEDs off),
the pneumatic switch will send a signal to the analyzer and the analyzer
will power up (if the Purge Bypass Switch is in the “
ACTIVE
” position, if
all fuses have been engaged, and if the external explosion-proof power-
disconnect switch has been closed).
A minimum of 15 minutes is required to properly purge and pressur-
ize the analyzer Electronics Enclosure.
The Expo Technologies MiniPurge
®
System uses a Battery Pack to
power its purge timer. AMETEK recommends replacing the Battery
Pack every three years to ensure proper and safe operation of the
purging system. See “Analyzer Preventive Maintenance Schedule” in
Chapter 6 for more information.
Working with the purged enclosure open is typically limited to installa-
tion, start-up, and certain troubleshooting and maintenance procedures.
In these cases, the Purge Bypass Switch (Figure 3-10) must be in the
“
BYPASS
” position (and appropriate safety conditions must have been
met, as per company policy).
For normal analyzer operation, the Purge Bypass Switch must be in the
“
ACTIVE
” position and the key must be removed (follow company policy).
!
WARNING
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...