Controller / User Interface | 4-17
RUN / CFG Mode – F2 Commands
Keystroke Term
Definition
F2
•
MBAddr
Modicon Modbus
®
slave address – can be any value between 0–255. ‘
0
’ disables the
Modbus
®
communication. Addresses 1–127 select the RS-232 port and addresses
128–255 select the RS-422 port.
F2 –
ComPara
zn
The setup parameters for the Baud Rate, Stop Bits, and Parity of the serial communication
port, where ‘
z
’ = 1..3 and where ‘
n
’ = 0..3.
n
= 0..3
Baud Rate
Stop Bits
Parity
z
= 1
z
= 2
z
= 3
0
9600
Invalid
None
1
4800
1
Odd
2
2400
1.5
None
3
1200
2
Even
Note:
Parameter changes must be saved to the EEPROM and the analyzer must be
reset for the changes to take effect.
To reset the analyzer, take all necessary safety precautions and then press
‘
•
’ and ‘
Ent
’ simultaneously.
F2 0
Cell
** Axial length of the Measuring Cell gas space (cm).
F2 1
Bench
n
** The Bench Type code, where ‘
n
’ = the current Bench Code value, defines which PMT is
designated as
Measure
and
Reference
, the number of temperature zones, and the filter
positions assigned to each source lamp.
Standard Bench Type Code
Code b6 (64) b5 (32) b4 (16) b3 (8)
b2 (4)
b1 (2)
b0 (1)
4 0 0 0 0 1 0 0
The value in (brackets) beside each bit number is its decimal weighting.
The typical value displayed is ‘
4
’ ((‘
4
’ = 3 temperature zones: 1, 2, 4). For systems with the
manual Blow-Back function enabled, the displayed value is ‘
2
’ (‘
2
’ = 2 temperature zones:
1, 4).
The extra temperature zones are not available when the Isolation and Blowback
functions are turned on.
See also “Probe Blow-Back Control” in this chapter.
Note:
The Standard Bench Type Code lists the default code for the analyzer.
Depending on the species being measured, the lamp types, and the application,
this code will change. For the Bench Type Code for your application, view the
Bench
screen (press
F2 1
in RUN mode).
Do not change the value without direction from AMETEK.
Bench Orientation (b0)
b0
Measure
Reference
0
PMT1
PMT2
1
PMT2
PMT1
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...