![Metal Samples MS2700E Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/metal-samples/ms2700e/ms2700e_operators-manual_1776978011.webp)
7
c.
Safe
Area
Installation
The
MS2700E
transmitter
is
approved
for
use
in
hazardous
areas,
but
can
be
used
in
non
‐
hazardous
areas
as
well.
For
basic
safe
area
wiring
information
refer
to
the
circuit
diagram
shown
on
page
21.
CAUTION:
When
used
in
non
‐
hazardous
areas,
equipment
must
be
supplied
with
a
pre
‐
approved
power
supply
unit
or
approved
equipment
which
meets
the
entity
parameters
shown
below.
The
pre
‐
approved
equipment
must
be
certified
to
the
electrical
safety
standards
for
equipment
in
ordinary
locations
(for
example,
EN/IEC/CSA
UL
61010
‐
1
or
EN/IEC/CSA
UL
60950
‐
1).
Failure
to
meet
these
requirements
will
void
the
intrinsic
safety
certification
of
the
equipment
and
prohibit
its
future
use
in
hazardous
areas.
d.
Hazardous
Area
Installation
CAUTION:
This
section
provides
general
guidelines
for
hazardous
area
wiring.
However,
regardless
of
anything
stated
here,
the
MS2700E
transmitter
must
be
installed
in
full
compliance
with
the
control
drawing
located
on
page
20
and
all
of
the
local
area
requirements.
CAUTION:
When
used
in
Hazardous
areas,
equipment
must
be
supplied
with
a
pre
‐
approved
power
supply
unit
or
approved
equipment
via
a
certified
intrinsically
safe
barrier
or
a
galvanically
isolated
barrier)
with
the
following
entity
parameters
.
Input supply terminal, RS-485 channel
and connector CN2 (Combined)
Connector CN1 pins A & B (J2 on PCB):
Ci = 68uF
Ui = 23 Vdc
Li = 0
Ii = 88.2 mA
Co = 32uF
Pi = 0.52 W
Lo = 0.14 mH
Ci = 0
Li
=
0
RS-485 channel at each line
Connector CN1 pins D & E (J9 on PCB):
Ui = 3.7 Vdc
Ii = 225 mA
Pi
=
0.206
W
U
o
= 5 V
I
o
= 51 mA
P
o
=
0.064W
Probe connector CN2 (J1 on PCB):
U
o
= 5 V
I
o
= 503 mA
P
o
= 0.7 W
Special
Conditions
1.
Probe
dielectric
rating
<
500V
r.m.s.
Do
not
exceed.
2.
Probe
cannot
be
installed
in
the
explosion
dust
atmosphere
(Group
III).
3.
Networked
(“daisy
‐
chained”)
arrangement
not
for
use
in
Group
IIC/IIIC
environment.
4.
Maximum
32
transmitters
are
allowed
in
the
daisy
‐
chain
arrangement.
Содержание MS2700E
Страница 2: ......
Страница 4: ......
Страница 23: ...19 Appendix A Drawings A Wiring Diagrams Control Drawing Hazardous Area Wiring Diagram Safe Area Wiring Diagram...
Страница 24: ......
Страница 26: ......