Using figures 11 and 12, connect the alarm module as follows:
1. Connect ribbon cable supplied with the alarm module from 25-pin connector on alarm
module to 25-pin "Alarm Out" connector (either 1 thru 16 or 17 thru 32) on rear panel of the
manifold controller.
2. Connect the alarm device between the applicable "+" voltage and the desired alarm number
output (1 to 16, or 17 to 32 if so connected).
3. Connect AC power cable from alarm module to facility power outlet.
Figure 12. Alarm Connector Plug Wiring
To acknowledge (turn off) a remote alarm, press the “Reset Alarms” key on the manifold controller.
On some older manifold controllers, it may be necessary to press, in order, “Local,” Reset Alarms,”
and “Remote.”