C-12-9800 Rev A
Page 35
Audit-Only Flow Checks
: see Section 7.4.2. No flow calibrations are changed if the ZERO and
CALIBRATE softkeys are not pressed.
7.4.2 Flow Audits
To audit any of the flow parameters, without making any calibration changes, enter the
corresponding calibration menu and allow the reading to stabilize. Verify the value is correct and
matches the flow standard. Record the results and press ESC to exit to the Calibration menu.
7.4.3 The CALIBRATE AUDIT Menu
– Performing Optical Span Tests
The CALIBRATE AUDIT menu is used to perform the optical span calibration verification test on
the instrument using a neutral density filter. This will introduce a known and consistent amount of
optical attenuation into the measurement system. The 83011 Span Check Filter tool will be
required to perform this test.
Note:
No changes to the instrument calibration are made with this process. This is a span check
only.
Press the MENU key. Highlight the CALIBRATE MENU item and select the ENTER key. Scroll to
the CALIBRATE AUDIT option and select the ENTER key again.
When entering the CALIBRATE AUDIT menu,
the Measurement Module should be in the
OPEN position. If it is not, or if the state is
unknown, press the MOVE button (see Figure
7-3) to open the measurement module and
advance the tape one position.
Figure 7-3 Entering the CALIBRATE AUDIT Menu
Press the START button to advance to the Zero measurement. Verify the filter tape is installed
and press the CONTINUE softkey. The measurement module will close, and the zero
measurement will begin as shown in Figure 7-4.
Figure 7-4 Performing the Zero Test
Once the zero test is complete, the span test must be performed using the 83011 neutral density
(ND) filter. The measurement module will open, and the screen will display instructions to insert
the ND Filter as shown in Figure 7-5.
VERIFY CALIBRATION
MOVE START