Page 10
C-12-9800 Rev A
3.5 Assembling the C-12
The C-12 is designed for easy setup and deployment. This allows it to be used for either
permanent long-term sampling at a fixed site or for portable audit sampling for temporary
applications.
Figure 3-1 The C-12 Shown with Optional 83598 Stand
The following describes the standard method of setting up the C-12.
1.
Insert the inlet tube:
Pull out the lanyard connected fitting from the inlet mount in the top
radiation shield opening. Insert the inlet tube.
2.
Connect the TSP Inlet to the inlet
: Slide the TSP inlet onto the top of the inlet. Be certain
the TSP inlet is fully seated onto the inlet.
3.
Apply power:
Plug in the AC power cable.
4.
Load filter tape
: Open the enclosure and load filter tape into the unit. Filter tape
installation is described in Section 3.7. Close the enclosure.
The C-12 monitor is now ready to begin sampling. See Section 3.8 for initial operation steps