4.4.1. Report Type
The Report setting determines the output format for the Serial Port. The choices are
NONE, CSV and PRINTER.
When set to NONE, the unit will not automatically output the reading at the end of a
sample to the Serial Port. CSV is a Comma Separated Values output format suitable
for importing into a spreadsheet. PRINTER is the same format as the screen and panel
mounted printer.
This setting does not affect the panel mounted printer which always prints in PRINTER
format.
4.4.2. Baud Rate
Use the Baud Rate selection to set the serial communications baud rate. BT-620
communicates at baud rates from 300 – 38400.
4.4.3. Serial Output Mode
The Serial Out setting controls the behavior of the BT-620 serial output. The modes
are RS232, RS485, Printer or Network (see Section 6 for serial communication
protocol). The following table lists the Serial Output settings and describes their
meanings.
Serial Out Setting
Description
RS232
RS232/USB communication.
RS485
RS485 communication.
NETWORK
RS485 communication with all serial output suppressed unless
specifically addressed.
4.4.4. Flow Control
The Flow Control setting is set to NONE for most standard RS-232 / USB serial port
applications. This setting can be set to RTS/CTS for hardware handshaking when using
the Ethernet port. The baud rate and flow control settings must also be set to match in
the Netburner Ethernet card setup for Ethernet connection.
4.5.
Printer Screen
Figure 17 shows the Printer Screen.
PRINTER: ENABLE