background image

Date:
19.11.13

Ux/14-100k

Page: A2

Manual - Appendix

Fig. a.2 shows the rear sides of transmitter(left) and receiver (right) with
connectors:

Signal connector SUB-D 25 (

Input: +/-15V 1M

Ω

||10pF

) and 

(

Output: +/-15V 10mA

)

The housing of the SUB-D connector is connected to the aluminum case,
which is connected to the circuit GND. This should be taken into account

during the test (possible ground loops, short circuits, parasitics to GND-
plane!).   The   pinning   is   shown   in  Fig.   a.2  and   is   also   printed   on   the

housings.

The receiver is also available as 19” rack mount kit (see  Fig. a.3), which

allows to integrate BNC plugs at the front (pinning: inner connector is
signal)   and   additional   other  plugs   in  parallel   (2mm   Hirschmann  in  the

shown case). Here, the 

SD-LED

 is integrated as well as the power switch.

At the rear side of the receiver, the connectors for optical input and the
power supply (6-8V, 0,5A) are placed. 

Notice, that the 19” rack mount kit is not made for use inside the EMC
chamber!

The SUB-D housing 
has the same 
potential as the 
aluminum case.

Fig. a.2: Rear side of the devices with connectors and switches

Input: 

+/-15V 1MΩ || 10pF

GND

GND

GND

GND

GND

GND

GND

GND

GND

16

14

12

10

8

6

2

4

13

15

11

9

7

5

3

1

PIN 1

Output: 

+/-15V max. 10mA

GND

GND

GND

GND

GND

GND

GND

GND

GND

2

4

6

8

10

12

16

14

3

1

5

7

9

11

13 15

PIN 1

Rear sides with 
connectors
Your pinning is 
printed on the 
housing!

Fig. a.3: front side of 19" receiver with connectors

Don't use 19“ rack 
mount receiver 
inside the ALC!

Distributed by: Reliant EMC LLC, 3311 Lewis Ave, Signal Hill CA 90755, 408­916­5750, www.reliantemc.com

 

Содержание U16/14-100k

Страница 1: ...Manual BV15 Analog transmitter U16 14 100k Distributed by Reliant EMC LLC 3311 Lewis Ave Signal Hill CA 90755 408 916 5750 www reliantemc com...

Страница 2: ...Distributed by Reliant EMC LLC 3311 Lewis Ave Signal Hill CA 90755 408 916 5750 www reliantemc com...

Страница 3: ...n 5 4 External filters 5 4 1 Differentiation input vs output filter 5 5 Maintenance 6 6 Trouble shooting 7 7 Accessories Options 9 8 Contact 9 Appendix Details and operation A1 Distributed by Reliant...

Страница 4: ...analog voltage signals Because of the optical transmission the system is very robust against EMS electromagnetic susceptibility It can withstand high electric and magnetic fields like they appear in...

Страница 5: ...o its input impedance do not use an extension Notice the differentiation of input and output filters see chapter4 1 4 1 Differentiation input vs output filter If you want to transmit analog voltages f...

Страница 6: ...rger to power the device inside the shielded room while EMI tests are running This might damage it Due to self discharge issues with NiMH batteries recharge batteries before use if the system has not...

Страница 7: ...at transmitter and receiver recharge batteries Test source signal directly at the device under test Device cannot be turned on cannot be charged Batteries damaged Internal fuse is broke Charger or ca...

Страница 8: ...r Comment Optical fiber LWL 1 xm x length in m simplex External batteries BP 60 25 6V 2 5Ah Connector cable for BP 60 SC 20 5m5m Length approx 20cm Charger with connector plugs CH1 5m Standard charger...

Страница 9: ...onnectors switches Fig a 1 shows the front sides of transmitter left and receiver right with connectors power push button with control LED Pwr On Off battery information LED Info charge plug Charge op...

Страница 10: ...el 2mm Hirschmann in the shown case Here the SD LED is integrated as well as the power switch At the rear side of the receiver the connectors for optical input and the power supply 6 8V 0 5A are place...

Страница 11: ...r into account while setting checking your measurement device Turn on all devices no order to be recognized Communication between transmitter and receiver will be indicated by the Signal detect LED at...

Страница 12: ...k messtechnik Others might lead to a damage of the system Only use battery packs and connector cables provided from mk messtechnik Other modules influence EMS performance and might damage the opto sys...

Отзывы: