Operating Instructions MBI 46.51.39
Version 0.9
en MBI 46.51.39 operating instructions.docx
Issue: 08/2022
page 13 from 29
5
Commissioning
WARNING
The amplifier may only be put into operation by qualified personnel.
Electrostatic discharges on electronic assemblies can lead to damage or direct failure of
components. Therefore, take all necessary measures to avoid electrostatic charging (ESD
protective measures).
Swapped or faulty connection, in particular external voltages connected to the signal output or
sensor input / output can destroy the device.
If the connection cable is laid properly, the distance between the displacement transducer and
the measuring amplifier can be up to 100 m and more.
Connect the inductive displacement transducer to the amplifier via a shielded, low-
capacitance cable.
Do not route the cable parallel to power lines and at a sufficient distance from electric
drives, transformers, and frequency converters that may create electromagnetic
fields.