V1.0/08-2018 | ENG

ECGSYS

WARNINGS

Do not place the electrodes on wounds, as this may cause additional injury.

It must be ensured that neither the patient nor the electrodes (including the neutral electrode) come 
into contact with other persons or conducting objects (even if these are earthed).

The MESI mTABLET ECG users must be adequately trained to use the device. Before the first use of 
the device, users must carefully read the entire instructions for use and follow the instructions for use 
of the connected equipment.

Do not use this unit in areas where there is any danger of explosion or the presence of flammable 
gases such as anaesthetic agents.

Protect the device from moisture and liquids and extremely high/low temperatures. Also, protect the 
device from mechanical stress and do not expose it to direct sunlight, as this can cause the device to 
not function properly.

MESI mTABLET ECG module must not be used in an oxygen-rich environment.

The ECG measurement unit is not intended for use in conjunction with high frequency surgical 
equipment.

Portable RF communications equipment including peripherals such as antenna cables and external 
antennas should be used no closer than 30 cm (12 inches) to any part of the device including cables 
specifed by the manufacturer. Otherwise degradation of the performance of this equipment could 
result.