12
Diagnostics
The Defender leak test is designed only to verify the
internal integrity of the instrument and alert you to an
internal leak. We recommend performing the Leak Test
only as an intermediate quality control check or when-
ever the integrity of the instrument is questioned due
to misuse or accidental damage.
Please note that a leak test is not a substitute for a
comprehensive examination of the unit’s overall per
-
formance and it does not ensure that your Defender is
operating accurately.
• Invert your Defender and allow the piston to travel to
the top.
• Cap the port under test using the supplied leak test
cap. Leave the other port uncapped.
• Press Enter on the control panel while the unit is still
inverted.
• Return the unit upright. The leak test will progress.