Q155 Service Manual
5-2
Figure 5-2 Air intake fine particle filter
5.4.
Bacteria Filter Replacement
The bacteria filter must be replaced after every 10,000 hours of use.
i.
Remove the cabinet and control panel.(See section 4.2.1)
ii.
Separate the tubing from both sides of the filter attached at control panel.
(See Figure 5-3)
iii.
Install the new filter with the inlet side in the same position as before. Push
the tubing together so that the tubing overlaps the bacteria filter connections.
iv.
Record information about the bacteria filter replacement on the History
Record card.
v.
Install the control panel and the cabinet.
Figure 5-3 Bacteria filter
Air intake fine
particle filter
Bacteria filter