Q155 Service Manual
4-2
4.3.
Unit Block Replacement
4.3.1.
Solenoid Valve Assembly Replacement
The oxygen concentrator uses solenoid valves to control airflow. It is not
necessary to do scheduled maintenance for the solenoid valves. If the valve has
got some trouble, you can easily replace the valve. The steps of replacing
solenoid valve are as following:
i.
Set the ON/OFF switch to the OFF position, and unplug the power cord.
ii.
Remove the cabinet. (See section 4.2.1)
iii.
Disconnect the terminal of the solenoid coil. (See Figure 3-8)
iv.
Remove the clamps and disconnect the tubes on the waste outlet/air
inlet/oxygen outlet, disconnect the PU tube from the pressure detect
fitting. (See Figure 4-3)
v.
Loose the six screws of solenoid valve and remove solenoid valve. (See
Figure 4-4)
vi.
Install new solenoid valve assembly and tighten six screws.
vii.
Reconnect the tubes and clamps of the waste outlet/air inlet/oxygen
outlet. Connect the PU tube to pressure inspect fitting located at valve
base.
viii.
Reconnect the terminal of solenoid valve.
Figure 4-3 waste outlet/air inlet/oxygen outlet
Air inlet
Oxygen outlet
Waste air outlet
Pressure detect
fitting