30
After the Zero Count Filter has been installed it is very important to clean (purge) the system before reading the values to qualify
the system.
Purge system
Press “Purge Time” to define the initial wait time and the purge time.
“Initial wait time” means the time delay before the pump starts working. For example, you can use this time to leave a clean room
and let the laminar air flow re-establish itself before starting the sampling.
“Purge Time” is the time when the pump is already working, but the counting has not started yet. The pump needs a time of about
10 seconds to reach a stable operating condition. Therefore, the purge time should not be set below 10 seconds.
setting Purge Timer
To purge the system first select the Purge Timer from the
main screen.
The screen will look like this briefly
This is the Purge Timer
Touch the screen here until
the Icon changes color
Содержание 1.44093.0001
Страница 1: ...APC SmartTouch Airborne Particle Counter Operation Manual Version 2 0...
Страница 21: ...21 Automatic method You are now connected to a LAN Network Now Press Here Now Press Here Now Press Here...
Страница 23: ...23 Type in the required IP Address and then press OK You are now connected to a LAN Network Now Press Here...
Страница 49: ...49 ISO 14644 range GMP range Average Min Max range...
Страница 56: ...56 The following screens show examples of each type FS 209 Memory ISO 14644 Memory GMP Memory...
Страница 57: ...57 Average Min Max memory As before all record types can be printed please see printing reports...
Страница 68: ...68 Press here Press here Then Press OK Press here...
Страница 69: ...69 Press here Then Press OK Press here Press here Then Press OK...
Страница 87: ...87...