3.1  Initial start-up

When the bath is started up for the first time, it should be supervised continuously until steady conditions 
have been reached.

3.2  Installation - waterbaths

Mains
connection 

The mains connection cable must be placed away from any hot surface.

 Filling

In order to garant the high quality of the stainless steel it is recommended by the 

steel producer to fill the waterbath with  deionised water.

 Draining 

The tank can be drained through the drain valve. The  thermostating fluid must be cooled 
to room temperature before  draining. Drain the fluid either directly into an appropriate 
container or connect an appropriate hose to the drain valve.

Operation with non-flammable thermostating liquids only!

drain valve

 

IMPORTANT! The water  level in the bath must be between the

two markings at the right side of the tub!

Waterbaths can be fitted with a

 level control system (see section 9).

NOTE:  Polluted fluids may not be drained directly into the public 

wastewater system. If the thermostating fluid is polluted, please 

make sure that it will be properly filtered or dumped according to the 

applicable regulations.

max water level

min water level

Содержание ONE 10

Страница 1: ...Waterbath WNE 7 45 WPE 45 Oilbath ONE 7 45 Operating Instructions...

Страница 2: ...EXCELLENT page 2...

Страница 3: ...monitor 15 4 3 7 Level monitor 15 4 3 8 Setup menu 15 4 3 8 1 Calibration 16 4 3 8 2 Selection of temperature monitor waterbaths only 17 5 Programming example of a programme sequence 17 6 Monitoring...

Страница 4: ...erating Instructions is necessary for faultless operation and for any possible claims under warranty If these Instructions are disregarded all claims under warranty guarantee and indemnification are e...

Страница 5: ...Weight kg 11 14 16 17 24 26 Current consume A 5 2 5 2 7 8 8 7 10 4 12 2 Power W 1200 1200 1800 2000 2400 2800 Ambient conditions Ambient temperature 5 C 40 C rh max 80 no condensation Overvoltage Cat...

Страница 6: ...C electromagnetic compatibility This product is intended to operate on a supply network with a system impedance Zmax at the transfer point building connection of 0 292 Ohm max The user has to ensure t...

Страница 7: ...1 4571 Class A in 4 wire circuit for control and monitoring Built in sounder as alarm on overlimit as audible signal at programme end and to acknowledge input key click Warning at low liquid level Co...

Страница 8: ...lve The thermostating fluid must be cooled to room temperature before draining Drain the fluid either directly into an appropriate container or connect an appropriate hose to the drain valve Operation...

Страница 9: ...rectly into an appropriate container or connect an appropriate hose to the drain valve For safety reasons the flash point of the oil must be at 300 C or higher The manufacturer of the oil may also hav...

Страница 10: ...e distribution 4 1 Controls and indications relatively largest heat loss since furthest away from the heated ribs minimum heat loss since very close to the heated ribs bath Insulation heating elements...

Страница 11: ...ol while holding down the SET key protection against unintentional alteration If the push turn control is rotated quickly the setpoint is altered in large steps with slow operation it is altered in si...

Страница 12: ...e maximum value until the display shows CCC The setting boiling mode is not stored permanently After the bath is switched off and on again the controller shows again the previously selected setpoint C...

Страница 13: ...flashes the hold time setting If no hold time has been programmed the hold time is activated immediately and the display alternates between the actual temperature and the running time of the hold tim...

Страница 14: ...ter circulation avoids larger temperature differences At a water temperature of e g 90 C the temperature difference in the entire bath is less than 0 36 C Note If the boiling mode CCC has been activat...

Страница 15: ...level has fallen so far that the upper of the two Pt100 temperature sensors is above the bath liquid the different temperatures inside and outside the bath liquid cause the symbol to flash the sounder...

Страница 16: ...e corrected 1 Set calibration temperature to 70 0 C and set calibration difference to 0 0 C 2 Using a calibrated reference instrument the measured temperature is 69 0 C at a setpoint temperature of 70...

Страница 17: ...nt to f ex 90 0 C C 2 Delayed switch on setting Rotate the push turn control until the delay and the symbol flashes With the SET key depressed use the push turn control to set the time f ex 6 00 C 3 H...

Страница 18: ...ctual temperature and SP until the setpoint temperature is reached Heating is indicated by the symbol C C Bath holds setpoint temperature SP is de activated or after the setpoint temperature has been...

Страница 19: ...after the TB cut out has been activated 1 Switch off the bath and allow it to cool down 2 Rectify the fault e g top up the liquid replace temperature probe and where appropriate contact customer serv...

Страница 20: ...the SET key The symbol flashes as warning 7 3 Monitor relay In addition the bath is equipped with an electronic monitor relay If a fault occurs during operation or if the selected setpoint temperatur...

Страница 21: ...erature of the thermofluid 9 Level control special equipment 2 different filling levels can be maintained constant if the unit is fitted with a level control system When using water as thermostating l...

Страница 22: ...at continuous influence can impair the outfit and function of the water oilbath Please use only detergents and antiliming agents appropriate for stainless steel for the cleaning of the tank and the h...

Страница 23: ...integrated digital timer for the hold time starts no earlier than the temperature in the bath has reached the setpoint to least 0 5 C at 95 C nominal temperature resp 2 C at 200 C nominal temperature...

Страница 24: ...2004 05 EN 61326 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amendments Council Direc...

Страница 25: ...t of domestic refuse For disposal please contact your dealer or the manufacturer Products which are infected infectious or contaminated with health endangering substances are excluded from return Plea...

Страница 26: ...8 L level control system 8 21 level monitor 15 liquid level oilbaths 9 liquid level waterbaths 8 low level protection 19 M maintenance 22 Mat Ref 1 4301 6 material quality 6 monitor relay 20 monitor...

Страница 27: ...EXCELLENT page 27...

Страница 28: ...Memmert GmbH Co KG Postfach 1720 D 91107 Schwabach Tel 49 0 9122 925 0 Fax 49 0 9122 145 85 E Mail sales memmert com www memmert com 24 04 2009 Water Oilbath EXCELLENT englisch D10333...

Отзывы: