Memmert ONE 10 Скачать руководство пользователя страница 6

EXCELLENT

page 6

2.1  Material quality

Memmert is using stainless steel ( Mat.Ref. 1.4301) for the external casing as well as for the interior, an 
outstanding material because of its high stability, optimum hygienic features and corrosion resistance 
against many (not all!) chemical combinations (Attention e.g. at chlorine combinations!).
The load has to be tested for its  chemical compatibility with the materials mentioned above.
A material-compatibility table covering all these materials can be requested from MEMMERT.

WARNING! Always pull out the supply plug before

opening the bath cover!

2.2  Electrical equipment

• 

Operating voltage see rating label 50/60 Hz, WPE 45: 50Hz

• 

Protection Class 1, i.e. operating isolation with ground connection to EN 61 010

• 

Protection IP20 to DIN EN 60 529

• 

Interference suppression to EN55011 Class B

• 

Bath protected by a fuse 250V/15A fast blow

• 

Controller protected by a 80 mA fuse (200 mA on 115 V)

• 

When connecting a MEMMERT bath to the electrical supply you have to observe any local    

 

regulations which apply (e.g. in Germany DIN VDE 0100 with FI protection circuit)

2.3 Note on EMC ( electromagnetic compatibility)
This product is intended to operate on a supply network with a system impedance Z

max

 at the transfer point 

(building connection) of 0.292 Ohm max. The user has to ensure that the product is only operated on an 
electrical supply network which meets these requirements. If necessary, details of the system impedance 
can be obtained from the local electricity supply authority.

Note:

Any work which involves opening up the bath must only be carried out 

by a properly qualified electrician!

2.4  Brief technical description

MEMMERT-water and oilbaths are electrically heated and electronically controlled.
The temperature of the  thermostating liquid is continuously controlled by a microprocessor-controller 
with pulse package control. Electronic microprocessor-PID-controller with continuous power matching 
and  autodiagnostic system with fault indicator (see section 12), integral timer for digital programme time 
selection.
The  temperature is measured using a  Pt100 temperature sensor (4-wire circuit). The accuracy is as 
follows:

WNE

ONE

Setting accuracy

0,1ºC

0,1ºC

 Temperature fluctuation

±0,1ºC

±0,2ºC

The components of temperature control are controlled by integrated malfunction-recognition. The heaters 
are installed outside - therefore no problems through dirt or limeresidue.

Содержание ONE 10

Страница 1: ...Waterbath WNE 7 45 WPE 45 Oilbath ONE 7 45 Operating Instructions...

Страница 2: ...EXCELLENT page 2...

Страница 3: ...monitor 15 4 3 7 Level monitor 15 4 3 8 Setup menu 15 4 3 8 1 Calibration 16 4 3 8 2 Selection of temperature monitor waterbaths only 17 5 Programming example of a programme sequence 17 6 Monitoring...

Страница 4: ...erating Instructions is necessary for faultless operation and for any possible claims under warranty If these Instructions are disregarded all claims under warranty guarantee and indemnification are e...

Страница 5: ...Weight kg 11 14 16 17 24 26 Current consume A 5 2 5 2 7 8 8 7 10 4 12 2 Power W 1200 1200 1800 2000 2400 2800 Ambient conditions Ambient temperature 5 C 40 C rh max 80 no condensation Overvoltage Cat...

Страница 6: ...C electromagnetic compatibility This product is intended to operate on a supply network with a system impedance Zmax at the transfer point building connection of 0 292 Ohm max The user has to ensure t...

Страница 7: ...1 4571 Class A in 4 wire circuit for control and monitoring Built in sounder as alarm on overlimit as audible signal at programme end and to acknowledge input key click Warning at low liquid level Co...

Страница 8: ...lve The thermostating fluid must be cooled to room temperature before draining Drain the fluid either directly into an appropriate container or connect an appropriate hose to the drain valve Operation...

Страница 9: ...rectly into an appropriate container or connect an appropriate hose to the drain valve For safety reasons the flash point of the oil must be at 300 C or higher The manufacturer of the oil may also hav...

Страница 10: ...e distribution 4 1 Controls and indications relatively largest heat loss since furthest away from the heated ribs minimum heat loss since very close to the heated ribs bath Insulation heating elements...

Страница 11: ...ol while holding down the SET key protection against unintentional alteration If the push turn control is rotated quickly the setpoint is altered in large steps with slow operation it is altered in si...

Страница 12: ...e maximum value until the display shows CCC The setting boiling mode is not stored permanently After the bath is switched off and on again the controller shows again the previously selected setpoint C...

Страница 13: ...flashes the hold time setting If no hold time has been programmed the hold time is activated immediately and the display alternates between the actual temperature and the running time of the hold tim...

Страница 14: ...ter circulation avoids larger temperature differences At a water temperature of e g 90 C the temperature difference in the entire bath is less than 0 36 C Note If the boiling mode CCC has been activat...

Страница 15: ...level has fallen so far that the upper of the two Pt100 temperature sensors is above the bath liquid the different temperatures inside and outside the bath liquid cause the symbol to flash the sounder...

Страница 16: ...e corrected 1 Set calibration temperature to 70 0 C and set calibration difference to 0 0 C 2 Using a calibrated reference instrument the measured temperature is 69 0 C at a setpoint temperature of 70...

Страница 17: ...nt to f ex 90 0 C C 2 Delayed switch on setting Rotate the push turn control until the delay and the symbol flashes With the SET key depressed use the push turn control to set the time f ex 6 00 C 3 H...

Страница 18: ...ctual temperature and SP until the setpoint temperature is reached Heating is indicated by the symbol C C Bath holds setpoint temperature SP is de activated or after the setpoint temperature has been...

Страница 19: ...after the TB cut out has been activated 1 Switch off the bath and allow it to cool down 2 Rectify the fault e g top up the liquid replace temperature probe and where appropriate contact customer serv...

Страница 20: ...the SET key The symbol flashes as warning 7 3 Monitor relay In addition the bath is equipped with an electronic monitor relay If a fault occurs during operation or if the selected setpoint temperatur...

Страница 21: ...erature of the thermofluid 9 Level control special equipment 2 different filling levels can be maintained constant if the unit is fitted with a level control system When using water as thermostating l...

Страница 22: ...at continuous influence can impair the outfit and function of the water oilbath Please use only detergents and antiliming agents appropriate for stainless steel for the cleaning of the tank and the h...

Страница 23: ...integrated digital timer for the hold time starts no earlier than the temperature in the bath has reached the setpoint to least 0 5 C at 95 C nominal temperature resp 2 C at 200 C nominal temperature...

Страница 24: ...2004 05 EN 61326 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amendments Council Direc...

Страница 25: ...t of domestic refuse For disposal please contact your dealer or the manufacturer Products which are infected infectious or contaminated with health endangering substances are excluded from return Plea...

Страница 26: ...8 L level control system 8 21 level monitor 15 liquid level oilbaths 9 liquid level waterbaths 8 low level protection 19 M maintenance 22 Mat Ref 1 4301 6 material quality 6 monitor relay 20 monitor...

Страница 27: ...EXCELLENT page 27...

Страница 28: ...Memmert GmbH Co KG Postfach 1720 D 91107 Schwabach Tel 49 0 9122 925 0 Fax 49 0 9122 145 85 E Mail sales memmert com www memmert com 24 04 2009 Water Oilbath EXCELLENT englisch D10333...

Отзывы: