INB 200 - 500

Operating instructiOns

Содержание INB200

Страница 1: ...Universal ovens UNB 100 500 UFB 400 500 Sterilisers SNB 100 400 SFB 400 500 Incubators INB 200 500 Operating Instructions...

Страница 2: ...BASIC page 2...

Страница 3: ...operation 10 5 1 Operating the door 10 5 2 Controls and indications 11 5 3 Switching on 11 5 4 Setting air changes 11 5 5 Setting the temperature 11 6 Selecting the operating mode 12 7 Normal operatio...

Страница 4: ...se Instructions are disregarded all claims under warranty guarantee and indemnification are excluded The right to technical modifications is reserved Dimensional details are not binding 2 1 Intended p...

Страница 5: ...our accessories 3 1 Subframe accessory Oven model 500 can be mounted on a subframe 3 2 Wall bracket accessory Oven models 200 to 500 can be wall mounted using the wall bracket The wall bracket is fac...

Страница 6: ...tion of inflammable gas air mixtures either within the oven chamber or in the immediate surroundings of the equipment Large amounts of dust or corrosive fumes inside the oven chamber or in the surroun...

Страница 7: ...500 800 900 Max number of trays 2 3 3 4 5 Max load per tray kg 10 30 30 30 30 Total load per oven kg 20 30 30 90 60 Ambient conditions Ambient temperature 5 C to 40 C rH 80 max no condensation Overvol...

Страница 8: ...re limiter TB Class 1 High grade PT100 temperature sensor Class A in 4 wire circuit Special equipment to be ordered separately as accessories subframe wall bracket sterilisation cassette 4 2 Material...

Страница 9: ...0100 with FI protection circuit This product is intended to operate on a supply network with a system impedance Zmax at the transfer point building connection of 0 292 Ohm max The user has to ensure...

Страница 10: ...gh ventilation slots 3 in the chamber side wall The fan 5 on the chamber back wall produces a large air throughput and a more intensive horizontal forced circulation compared with natural convection T...

Страница 11: ...ET key and set the temperature setpoint with the push turn control After the SET key has been released the display briefly flashes the temperature setpoint The display then changes to the actual curre...

Страница 12: ...l operation In this operating mode the oven operates continuously and heats and controls to the set temperature On the ovens UFB SFB the fan is running continuously Setting the temperature Hold down t...

Страница 13: ...oven will remain switched on Setting the temperature Turn the push turn control clockwise until the temperature display is flashing Hold down the SET key and set the required temperature setpoint with...

Страница 14: ...ter it has cooled down and after the fault has been rectified 9 2 Monitor relay In addition to mechanical temperature protection the oven is provided with an electronic monitor relay If a fault occurs...

Страница 15: ...cabinet The operation of the MEMMERT hot air steriliser is also subject to the Standard DIN 58 947 Part 6 A few examples of the correct preparation for different medical products are summarised in th...

Страница 16: ...e 180 C heating up time sterilisation time total sterilisation time Type of loading light medium heavy Steriliser size withoutfan withfan withoutfan withfan withoutfan withfan 100 0 45 h 0 45 h 1 30 h...

Страница 17: ...tes The cassettes should preferably be so arranged in the steriliser that the hot air flow can pass readily through the air slots The load to be sterilised is placed into the steriliser cassettes wrap...

Страница 18: ...bly qualified electrician MEMMERT products require little maintenance It is however recommended to lubricate all moving parts of the doors hinges and closure once a year or 4 times a year with continu...

Страница 19: ...module faulty E 3 PT100 temperature probe faulty In case there is a fault on the oven please get in touch with an authorised service organisation or contact the MEMMERT customer service department Whe...

Страница 20: ...ive DIN EN 61326 2004 05 EN 61326 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amendmen...

Страница 21: ...6 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amendments Council Directive on the appr...

Страница 22: ...regulations of the directive 93 42 EEC Directive of the council to adapt legal regulations of the member states on the subject of medical products dd 14 06 1993 Abl EG Nr L 169 S 1 12 07 1993 includin...

Страница 23: ...al Law It should not be disposed off as part of domestic refuse For disposal please contact your dealer or the manufacturer Products which are infected infectious or contaminated with health endangeri...

Страница 24: ...5 M main switch 11 maintenance 18 material quality 4 7 medical product 4 monitor relay 14 N normal operation 12 O operating mode indication 11 operating mode selection 12 operating time 13 operation...

Страница 25: ...BASIC page 25 Notes...

Страница 26: ...BASIC page 26 Notes...

Страница 27: ...BASIC page 27...

Страница 28: ...Memmert GmbH Co KG Postfach 1720 D 91107 Schwabach Tel 49 0 9122 925 0 Fax 49 0 9122 145 85 E Mail service memmert com www memmert com 12 08 2009 BASIC english D09803...