ICO/ICOmed/HCP stacking
D39166 | date 09/2019
3.4 Fitting wall brackets
NOTICE
The adjustable wall brackets from the stacking set must be used for size 150 and 240
appliances.
1. Remove one M3x6 Torx screw from each of the appliances on the left or right side of
the frame.
2. Fit the wall brackets to the rear frame.
3. Drill holes, insert dowels and screw the tilt protection brackets into a suitable wall.
►
Since the conditions (wall structure) at the installation site are unknown, no dowels
or similar materials for wall mounting are included in the scope of delivery.
The installation is now completed.