Appendix D:
Detector Calibration
This procedure performs the aligment of the SARK-110 internal measurement
circuits. This operation is done once at the factory and it is not necessary that the
user do this operation but this is provided anyhow.
For the calibration it is required the following test loads: 0, 50, 100, and 200-ohm.
These loads need to have good precission and stability and should be properly
implemented in order to have the minimum stray impedance.
Before performing the operation it is highly recommended backing-up the
“detcalib.dat” file that contains the factory adjustments.
The calibration operation is available in the «Setup» «Calibration» «Detector
Calibration».
The calibration procedure is as follow: