background image

 

 

 

We reserve the right to make technical modifications without prior notice. 

 

User Manual

 

 
 

for the 

 

Vacuklav

 31

 

Autoclave 

 
 

 

Dear Doctor: 

Thank you very much for the trust which you have shown by purchasing this autoclave. 

For nearly 50 years now, MELAG — a medium-sized family-owned and -operated busi-
ness — has specialized in the production of sterilization equipment for medical practice.  
During this period, MELAG has succeeded in becoming a leading manufacturer of sterili-
zation equipment. More than 300,000 MELAG units sold throughout the world testify to 
the exceptional quality of our sterilizers. 

As all other MELAG products, this autoclave was manufactured and tested according to 
strict quality criteria. Before placing this unit into operation, please carefully read this User 
Manual. The long-term functional effectiveness and the preservation of the value of your 
autoclave will primarily depend on careful preparation of instruments before sterilization, 
and on proper care of the unit. 

The staff and management of MELAG 
 

 
 

 

Содержание Vacuklav 31

Страница 1: ...ucceeded in becoming a leading manufacturer of sterili zation equipment More than 300 000 MELAG units sold throughout the world testify to the exceptional quality of our sterilizers As all other MELAG...

Страница 2: ...nitoring 10 2 4 2 Periodic testing every six months 10 2 4 3 Maintenance service of the autoclave 10 3 Installation and initial startup 11 3 1 Setting up the autoclave 11 3 1 1 Transport ribbons 11 3...

Страница 3: ...ransport placing the autoclave back into operation 26 6 Malfunctions 27 6 1 Malfunctions without error display 27 6 1 1 Display of the operational state 27 6 1 2 A program does not start 27 6 2 Malfun...

Страница 4: ...r gasket 3 Display 11 Twist door handle 4 Front panel with front film scale 12 Door locking nut 5 Enclosure cover above the water storage tank 13 Door 6 Door locking spindle 14 Door contact 7 Housing...

Страница 5: ...error 36 Arrow pointing to power switch 1 1 3 View of rear _________________________________________________________________________ 38 Sterile filter 74 Data interface for printer 39 Connection for w...

Страница 6: ...into the program sequence It ensures excellent drying results within a short drying period This optional supplementary drying enables the operator to extend the standard drying time in order to achie...

Страница 7: ...ruments to be sterilized are thoroughly cleaned before sterilization in order to prevent contamination from being separated from the instruments during sterilization and from clog ging the filters and...

Страница 8: ...storage tank for example by using a small brush with hot water and a grease cutting additive Or use a non raveling cloth soaked in alcohol Then thoroughly rinse out the water storage tank with water a...

Страница 9: ...iO2 0 1 mg kg 1 mg l Iron 0 1 mg kg 0 2 mg l Cadmium 0 005 mg kg 0 005 mg l Lead 0 05 mg kg 0 05 mg l Rest of heavy metals ex cept iron cadium and lead 0 1 mg kg 0 1 mg l Chlorides Cl 0 1 mg kg 2 mg l...

Страница 10: ...m the results of sterilization on an official test log Spore tests to confirm the effectiveness of sterilization of an autoclave provide valid results only in conjunc tion with the sterilization progr...

Страница 11: ...the autoclave is to use a spirit level carpenter s level to cor rectly align the unit The spirit level must first be held up flush against the flange of the chamber to ensure that the entire unit is v...

Страница 12: ...onnected to the condensate container 80 43 One way drain system 73 Connection for drain to con densate container 80 Condensate container 81 Connection for drain from autoclave 82 One way drain system...

Страница 13: ...rts 23 Then start the de sired program again 3 3 2 Option connection from condensate container to a building drain___________________________ An option is automatic draining of the condensate containe...

Страница 14: ...the water storage tank over the maximum filling mark MAX 3 5 2 Automatic water refill_________________________________________________________________ As delivered by the manufacturer the Vacuklav 31...

Страница 15: ...water as supply into the MELAdem 45 reverse osmosis system drain valve with pipe vent non return valve 1 2 x 3 4 Drain water connection fitting for drain water outflow from the MELAdem 45 reverse osmo...

Страница 16: ...ng feed hose from the MELAdem 45 reverse osmosis system to the threaded fitting of the control unit for external water refilling at the external solenoid valve Item 2 Wastewater connection Use the uni...

Страница 17: ...eaks Test for overflow of the autoclave This test is carried out by blocking the float switch in the water storage tank of the autoclave The user must then check to ensure that the inflowing water is...

Страница 18: ...ece in 01045 For more detailed information on the connection of water treatment systems by other equipment manufac turers please consult your specialist MELAG dealer and mention the key words External...

Страница 19: ...time the autoclave system will balance itself The test itself lasts ten 10 minutes If the system determines a leakage rate of 1 3 mbar min the test is considered successful and the follow ing completi...

Страница 20: ...n Program select 18 When you press start stop 19 you have installed the printer 3 9 VDE stipulations ________________________________________________________ According to VDE stipulations presently va...

Страница 21: ...amage to the autoclave and to the instruments being sterilized 4 4 Loading the autoclave ___________________________________________________ The operator can use either trays containers or standard tr...

Страница 22: ...lave chamber The display will show the progress of the fractional evacuation as follows Display line 1 Sub atm pulsing Display line 2 0 2 bar 60 C Fractional evacuation will be repeated several times...

Страница 23: ...C Display line 1 Immediat removal Display line 2 Press STOP 4 13 Supplementary drying __________________________________________________ The Vacuklav 31 has been provided with the supplementary drying...

Страница 24: ...ing are various print and data saving possibilities which can be initiated by pressing the button for printing 22 and the and buttons 17 Pressing the button for printing once Display line 1 Immed prin...

Страница 25: ...release The display then notifies the user of the type of mal function In addition if a printer is connected to the autoclave it will print out the malfunction report and will also show the values fo...

Страница 26: ...t and allow the autoclave to cool down Remove the hoses and tubes from the rear of the autoclave Open the tap for draining the water storage tank and allow the water to run out of the tank If you wish...

Страница 27: ...cket carrying the proper line voltage If the plug is in the socket and if the power switch is on the autoclave display will light up If line voltage is provided and if the power switch is on but if th...

Страница 28: ...ralized water aqua dem aqua dest Press the button for acknowledg ing error reports 23 Then start the program again MALFUNCTION Temp too high The malfunc tion signal lamp lights up 35 The steam tempera...

Страница 29: ...cartridge module The malfunction signal lamp lights up 35 The conductivity of the water is too high You can still start your program however by pressing the start stop button 19 Empty the water stora...

Страница 30: ...ed instruments 19 24 15 min 10 min Textiles 33 41 15 min 10 min Disinfection 20 27 15 min 10 min Bowie Dick Test approx 26 min 5 min 5 min Vacuum test approx 16 min 7 1 3 Electrical connections consum...

Страница 31: ...ys Max of 1 5 kg of textiles Possibilities of loading Tray racks with a max of five 5 MELAG trays Tray racks with a maximum of three 3 standard tray cassettes or three 3 MELAG trays MELAG sterilizatio...

Страница 32: ...heating these objects and the containers As a result of this process the steam undergoes a change of state from gaseous to liquid This liquid conse quently appears as condensate on the instruments and...

Страница 33: ...acked vertically fig B This vertical configuration ensures that channels can form between the textile folds for the air to flow out and for the steam to flow in Do not stack textiles horizontally fig...

Страница 34: ...agents not only protect the instruments themselves they also protect the microorganisms on the instruments 8 4 Loading the autoclave ___________________________________________________ Do not sterili...

Страница 35: ...ely to the bottom of the chamber The best solution arrange several sterilization containers of the same size one on top of each other so that the condensate can run down the side walls of the steriliz...

Страница 36: ...r evaporate the rest of the condensate after the sterilization process has been completed German Standard DIN 58953 Part 7 Section 7 contains the following information concerning moisture which may re...

Отзывы: