5 Sealing
15
3.
Leave the sealing lever depressed as long as the
progress bar in the display completes and the control
lamp flashes green at short intervals (approx. 3 s).
4.
If the display switches to
Open
and the control lamp is
continuously illuminated green, return the sealing
lever upwards to its starting position.
Ê
If the USB flash drive has been connected, a log
will automatically be generated and saved on
the USB flash drive.
Ê
If a USB flash drive has not been connected, the
most important sealing parameters will be
displayed alternately up to four times. Then the
display changes to
Ready
.
5.
Perform a visual inspection of the film after every
successful sealing procedure.
PLEASE NOTE
With bags with side gussets, perform weekly controls of the seal seam using an ink test
(e.g. MELAink test).
Sealing procedure for film rolls
Produce film bag
To package instruments in transparent sterilization packaging bags from the roll, proceed as follows:
1.
Guide the end of the film roll along the device rear
panel into the lower paper guide with the paper side
facing downwards (
lower slit
). Slide the film forwards
and between the pressure and sealing rail to the
desired length.