3 Description of the device
11
Status display and acoustic signals
Control lamp /
acoustic signal
Possible cause
Measure
LED illuminates orange
The sealing device is in the
heating or cooling phase.
Wait until the pre-set sealing temperature
has been reached.
LED flashes red,
warning signal sounds
The sealing lever is depressed
during the heating phase.
The sealing temperature has not
yet been reached.
The sealing lever has been raised
early, despite the required sealing
duration not having been reached.
The sealing lever has not been
raised, despite the required
sealing duration having been
completed.
Acknowledge the display message with
OK
and wait until the LED control lamp is
continuously illuminated green.
Acknowledge the display message with
OK
and leave the sealing lever depressed
during sealing until the message
Open
is
issued.
Raise the sealing lever as soon as the
display message
Open
is issued, otherwise
the sealing procedure is unsuccessful.
Failure to comply with these provisions can
result in the melting of the film and damage
to the sealing device.
LED illuminates green
The sealing device has reached
the pre-set sealing temperature
and is ready for operation.
The minimum sealing duration has
been reached and the sealing
procedure has been ended.
Raise the sealing lever and remove the
packaging.
LED flashes green
The sealing procedure runs when
the sealing lever is depressed; the
progress bar is displayed.
Wait until the green LED control lamp is
continuously illuminated or the display
shows
Open
.
LED illuminates red,
warning signal sounds
(malfunction)
Device malfunction: The heating
phase is taking too long (> 5 min);
the sealing device does not reach
the pre-set sealing temperature.
Switch off the sealing device and then on
again. Upon repeated incidence, contact an
authorized customer services or stockist
technician.