Chapter 4 – Malfunctions
10
Chapter 4 – Malfunctions
Incident
Possible cause
What you can do
No distilled water is produced
The mains plug is not in the
mains socket
Mains socket is defective
Upper section connector cable is
not on the distilling chamber
START/RESET button is not
depressed
Check the mains plug
Check the electricity supply on
the mains socket
Connect the connection cable to
the distilling chamber
Press the START/RESET button
Too little distilled water produced Upper section connection cable is
not in the distilling chamber
Distilling chamber is not filled
sufficiently
The cover seal does not seal
sufficiently, water escapes
Glass flask positioned incorrectly
under the water distiller
Cooling system soiled
Connect the connection cable to
the distilling chamber
Refill with tap water
Replace the cover seal in the
upper section of the distilling
chamber
Check whether water flows past
the glass flask and place the
glass flask under the water
distiller correctly
Clean the cooling system, blow
out with compressed air if
necessary
Steam escapes under the
discharge opening of the upper
section
Upper section connection cable is
not in the distilling chamber
Connect the connection cable to
the distilling chamber
The water distiller does not heat The water distiller overheats
Allow the water distiller to cool for
5-10 minutes, then restart
The water distiller does not
deactivate automatically
Overheating switch defective