Chapter 4 – Maintenance
13
8.
Switch on the steam sterilizer and open the pressure tank shut-off valve.
9.
Check whether all the parts are secure and that the reverse osmosis unit is free from leaks.
Replacing the activated carbon filter
Replace the cartridge of the activated carbon filter once a year or when replacing the fine filter.
1.
Switch off the steam sterilizer and close the
pressure tank shut-off valve.
2.
Unscrew the filter housing (a) from the filter
body (b) using the filter key. Guide the filter
key upwards from below over the filter
housing and turn the filter key in a clockwise
direction.
WARNING
The filter housing is full to the brim with
water. Remove the water.
3.
Remove the activated carbon filter (1) and
rinse the filter housing with tap water.
4.
Remove the sealing ring from the filter housing (a).
5.
Clean and grease the sealing ring lightly with silicone grease (art. no. 24371).
6.
Insert the new activated carbon filter in the filter housing.
7.
Screw on the filter housing lightly using the filter key.
8.
Switch on the steam sterilizer and open the pressure tank shut-off valve.
9.
Check whether all the parts are secure and that the reverse osmosis unit is free from leaks.
b
1
a