3 Carebox
7
Servicing
The following maintenance activities are described in the Careclave user manual.
Interval
Measure
Device component
Daily
Check the Carebox for contamination
and clean if necessary
Carebox
Check the Carebox sieve for
contamination and clean if necessary
Carebox lower section
Check O-rings on the adapters for
presence, wear or damage and replace if
necessary
Carebox upper section
Visual inspection of the seals (housing
seal and media seals) and replace if
necessary
Carebox
Weekly
Cleaning
Carebox
After 600 cycles
(recommended)
Change adapter seals (O-rings)
Carebox upper section
Spare parts
Device component
Article
Art. no.
Carebox lower section
Carebox filter (12 pcs.)
ME21412
Carebox lower section
Housing seal
ME21404
Carebox upper section and
Carebox lower section
Set media seals for Carebox:
2x media seal - big
5x media seal - small
ME21465
Carebox lower section
Sieve for Carebox
ME10701