Meilhaus Electronic Red-USB Iso Скачать руководство пользователя страница 4

USB-Isolator U

ser’s Manual              

                

_________________________________________________________________________________ 

 

_______________________________________________________________________________________ 
 

 

1)  Connect B-type plug 

(square connector) to ‘PC connect’ side of the USB 

Isolator. 

 

 

 

2)  Connect A-type plug of USB cable to an available USB port on the 

computer (the power LED will turn on when connected to a powered 
computer). 

 

 

 

3)  Connect the A-type plug 

(flat connector) to the ‘device connect’ side of 

the isolator.  Connect the other side of the cable to the peripheral being 
isolated.  

NOTE:  

This device is not intended to isolate the trunk of a 

USB hub, as its output current is limited to ~137mA.   

 

 

 

4)  Your computer should prompt that a new device has been connected to 

the computer.  The power LED of the isolator should be ON and not 
blinking.  At this point, you are ready to begin talking to your connected 
device. 

 
TIP:  

If your PC fails to detect your USB device, try pushing the reset 

connection button on the Isolator.

Installation 

 
Guide to Quickly 
Establishing 
Communication 
 

Note:  

This device is not 

intended to isolate hubs. 

 
Note:  

This isolator will break 

the path to earth ground.  Most 
personal computers connect 
the USB signal ground in 
common with the cable shield 
and earth ground.  The PC 
connector side of this isolator 
also holds signal ground in 
common with the connector 
shield.  The target device 
connector side of this isolator 
does not transfer the earth 
ground connection across the 
barrier that was made at the 
PC side.  If the device that you 
are isolating must be earth 
grounded, other provisions 
must be implemented to 
introduce earth ground to the 
isolated side. 

 
 
 
 
 
 
 
 

Содержание Red-USB Iso

Страница 1: ...Red USB Iso USB Isolator User s Manual Ausgabe May 2011...

Страница 2: ...er USB cable Type A to Type B USB 2 0 compatible Changeable Data Rate allows selectable full speed 12Mbps Default or low speed 1 5Mbps Wide Ambient Operation from 40 to 70 C 40 to 158 F Hardened For H...

Страница 3: ...esponding to 137mA out load Both signal lines are equipped with common mode filters and include transient protection USB is set to full speed 12 Mbps but can be jumped to low speed 1 5 Mbps for legacy...

Страница 4: ...ice has been connected to the computer The power LED of the isolator should be ON and not blinking At this point you are ready to begin talking to your connected device TIP If your PC fails to detect...

Страница 5: ...closure and this requires handling at an ESD safe work station Warning We do not recommend that you do this in the field as it invites potential damage to sensitive internal circuitry 1 Open enclosure...

Страница 6: ...ot communicate with device connected through the Isolator Your computer does not support the peripheral Your device does not support 12Mbps Try changing internal jumpers to low speed see page 5 You do...

Страница 7: ...discharge to the enclosure port per IEC61000 4 2 Radiated Field Immunity RFI 10V M 80 to 1000MHz AM and 900MHz keyed carrier per IEC61000 4 3 Electrical Fast Transient Immunity EFT 2KV to power and 1K...

Страница 8: ...hotocopying recording or otherwise without the prior written permission of Meilhaus Electronic GmbH All the information included in this user s gide were put together with utmost care and to best know...

Отзывы: