47 / 48
9617722
Note to customers
Dishwashing machines, bedpan rinsing units and systems
are intended for fixed, electrical power supply installation as
well as connection to the locally available equipotential
bonding and have been equipped with a corresponding con-
nection option.
Operators may decide at their own discretion and responsi-
bility to implement personal protection in locally available
services in collaboration with a specialist electrician regis-
tered at the corresponding energy supplier using the follow-
ing:
AC/DC sensitive fault current protection switch with
at max. 30mA EN 62423
or
Automatic shutdown of the supply in the event of loss
of protective earth conductor conductivity (EN 60204-1,
Section 8.2.8.c)
12.1
EC/EU Declaration of Conformity
see separate EC/EU Declaration of Conformity
13. Further documents
According to the authorization level, additional documents are available for this operating
instructions:
Installation instructions
Service instructions (only for authorised service technician or house engineer)
Short operating instructions
Specific list of suitable care utensils
Installation and operation qualification