FOCUSING PROCEDURE

The whole point of stereoscopic microscopes is that, when properly adjusted, the observer sees the object 
image in three dimensions and can perceive height and depths on the specimen.  This is a great advantage 
in material examination and quality control techniques.

To achieve these desirable height/depth effects the images coming from the binocular eyepieces must be 
"fused" into a single image by the observer - this requires some practice and careful setting of the binocular 
body.  Procedure is as follows:-

The operator should move eyepiece tubes in and out to find the place where the distance between 
eyepiece centers matches the distance between the pupils of his own eyes.  This is the 

interpupillary 

distance 

and will vary somewhat from operator to operator.  When these distances are equal (or 

match

), 

one central image is seen by the operator.

54-74mm

"Fused"

When replacing stage plate loosen the set screw 
which is visible infront of the stage.

Stage plate set screw

Black and White
Stage Plate

The Rigid arm type focusing stands (except the 
model ABE) have a focusing capability only by 
rack and pinion adjustment.  This means that the 
available range of specimen sizes/depths which 
can be accommodated is somewhat less than in 
the case of Pole-type stands.

The Rigid arm type ABE focusing stand is with 
integral focus block and extendable working 
distance.  Provides transmitted and incident 
halogen illumination.

To extend working distance, 
loosen the clamp screw and 
extend the arm upward and 
set the screw back.

AZ, ABZ and ABZH 
stands are rigid arm type
focusing stands.

Focusing Screw

Clamp Screw

ABE Extendable

PRECISE FOCUS ON SPECIMEN

Stage Plates

You should now decide which of the stage plates supplied will be most suitable.

The focusing stands with plain bases, P, PL, PX, PC and A, are supplied with reversible black and white 
stage plates. Your selection depends on which side, black or white, gives the best contrast conditions with 
your selected specimen.

The focusing stands with transmitted illuminators, PB, PBH, ABZ, ABZH and ABE, are supplied with clear 
glass stage plates and reversible black and white stage plates.  If the specimen is semi-transparent, the 
glass stage plate shoud be used with transmitted illumination switched on.  If the specimen is opaque, use 
the black/white stage.  Your selection depends on which side, 
black and white, gives the best contrast.

Содержание EMZ-2

Страница 1: ......

Страница 2: ...How to Operate Illuminators Focusing Capability Precise Focus On Specimen Focusing Procedure PHOTOGRAPHY AND TELEVISION Photography Camera Operation Television MAINTENANCE AND CARE Bulb Replacement Care Cleaning SYSTEM DIAGRAM 2 3 4 5 7 5 5 6 6 6 7 7 8 7 8 8 9 10 13 1 CONTENTS EMZ Series INSTRUCTION MANUAL ...

Страница 3: ...Clamp Collar Set Screw Pillar Collar Dimmer Switch Eyepieces with Eyeshield Eyepieces with Eyeshield Eyepieces with Eyeshield Diopter Adjustment Ring Diopter Adjustment Ring Zoom Control Knob Diopter Adjustment Ring Zoom Control Knob Objectives Incident Illuminator Lamp Cover Stage Clips Stage Plate 2 EMZ 5TR Trinocular Zoom Stereo Body Eyetube Inclination 45 EMZ 8TR Trinocular Zoom Stereo Body Ey...

Страница 4: ...Zoom Control Ring Clamp Screw Set Screw Model EMZ 2 with SWF10X mounted on P Stand 3 EMZ 9 Zoom Stereo Body Eyetube Inclination 60 Z 7100 Zoom Stereo Body Eyetube Inclination 90 ...

Страница 5: ...h gives firm and stable support This should be located where the atmosphere is as clean as possible avoiding places where there is excessive dust moisture heat or fumes ASSEMBLY When in place insert eyepieces in the eyetubes of the binocular body Zoom stereo body are already in place on the limb as shipped Cover with the plastic dust cover supplied until ready for adjustment and use If you ordered...

Страница 6: ...Select an illuminator to use either incident or transmitted by the selector switch in the way described above Then turn the dimmer switch knob slowly to clockwise to turn the dimmer switch ON and to increase brightness Turn the knob to counterclockwise to decrease the brightness and to turn OFF OFF I T I T OFF Selector Switch Focusing Knob Focusing assembly Clamp Screw Pillar collar and clamp Scre...

Страница 7: ...ability only by rack and pinion adjustment This means that the available range of specimen sizes depths which can be accommodated is somewhat less than in the case of Pole type stands The Rigid arm type ABE focusing stand is with integral focus block and extendable working distance Provides transmitted and incident halogen illumination To extend working distance loosen the clamp screw and extend t...

Страница 8: ...cular photo binocular bodies offered for use with MEIJI TECHNO microscpoes In the case of the EMZ TR Zoom Stereo series a trinocular is supplied with lever switching of the image from one of the binocular eyepieces vertically up to the film plane of a 35mm SLR camera with adaptor mounted on the vertical photo tube Visual observation can be simultaneously carried out using the other eyepiece In thi...

Страница 9: ... camara and the visual eyepiece Rotate the adjustment ring on the straight tube so as to set the red indicator line on the upper tube portion at the top edge of the lower tube Then fix in place with the clamp screw You now should be set correctly for optimum conditions of simultaneous visual observation and photography CK3900N or CK3900P CCD Camera MK300N or MK300P B W CCD Camera CK3900N or CK3900...

Страница 10: ...BZH ABEstand are same halogen Catalog number is MA570 CARE Always cover the instrument with the plastic dust cover provided when the microscope is not in use Keep the eyepieces in the microscope body at all times in order to prevent dust from falling on the internal optics Store the microscope in a safe clean place when not in use for an extended period of time CLEANING Clean exposed lens surfaces...

Страница 11: ...er PBH Pole type w halogen 6V 10W trans incident PB Pole type w tungsten 6V 7W trans incident lights PC Plain Pole w coaxial focusing PL Pole type w illuminator port MA568 B W Plastic stage plate MA551 Extension pole PX Pole type w illuminator port 10 Piller P Plain Pole type AZ Plain Rigid arm MA565 Ungraduated mechanical stage MA564 Graduated mechanical stage MA531 Protective glass MA550 Polariz...

Страница 12: ...W trans incident lights ABE Extendable w halogen 6V 10W trans incident lights BX Stand w mirror for B D light not included MA501 SWF5X MA502 SWF10X MA520 SWF12 5X MA503 SWF15X MA535 HSWF15X MA504 SWF20X MA521 SWF30X MA519 SWF10X F Piece MA600 EYESHIELD MA564 05 Graduated mechanical stage with clear glass plate MA565 05 Ungraduated mechanical stage with clear glass plate EYEPIECES Paired MA545 0 3X...

Страница 13: ... Stand UL Universal Stand MU Universal Stand MA574 Large Sliding Stage MA575 Medium Sliding Stage MA574 Large Sliding Stage MA575 Medium Sliding Stage FA 1 Articulated Arm Stand FA 2 Articulated Arm Stand EMF 1 EMF 2 EMT 1 EMT 2 EMT 3 EMT 4 EMTR 1 EMTR 2 EMTR 3 EMTR 4 EMF EMT VIEWING HEADS EMZ 1 EMZ 2 EMZ 5 EMZ 5D EMZ 6 EMZ 9 EMZ 10 Z 7100 EMZ 2TR EMZ 5TR EMZ 5TRD EMZ 8TR EMZ 8TRD ZOOM HEADS EMF 1...

Страница 14: ...nt C Mount MA151 10 C Mount with lens C Mount with lens MA151 8TR 0 6X C Mount MA151 10 C Mount with lens MA151 8TR 0 6X EMZ 8TR EMZ 8TRD EMTR 1 EMTR 2 EMTR 3 EMTR 4 EMZ Series TRINOCULAR VIEWING HEADS EMT VIEWING HEADS TRINOCULAR EMZ 5TR EMZ 5TRD EMZ 2TR EMZ 8TR EMZ 8TRD EMTR 1 EMTR 2 EMTR 3 EMTR 4 EMZ Series TRINOCULAR VIEWING HEADS EMT VIEWING HEADS TRINOCULAR EMZ 5TR EMZ 5TRD EMZ 2TR EMZ 8TR E...

Страница 15: ......

Отзывы: