HT 7330100000
INFRARED THERMOMETER KC-180B-1
Page - 3
6
Press the middle key for 2s, then short press the left down key and the right down key to set the emissivity, low and
high temperature threshold. Finally hold the middle key to finish the setting.
7
When the battery is low, the low battery icon “” will appear on the upper area of the screen. The “” and “” will appear
on the upper of the display if the reading is lower or higher than the pre-set low(LAL) and high(HAL) set.
8
While inactive status, press the right upper key to switch between infrared measuring and thermocouple measuring.
When appears, then plug in the type K-thermocouple into the socket which located on the right side of the instrument.
Pull the trigger to get the reading.
Operation Introductions
1. Point the product to target objects, hold the trigger to get the reading.
2. The distance and scanned area: the value of D:S must be considered ( as showed ), the ratio should be no more
than 12:1. In other words, if you are 12 inch(300mm) from the target, the detector will measure the average
temperature over a circle 1 inch(25mm) in diameter. The field of view should cover the surface of measured objects.
3
. Field of view: Be sure the target area is larger than the unit’s spot. The smaller the target area, the closer the
distance between unit and target. However, if you want to have an accurate temperature of an object, be sure that
the target is twice as large as the measuring area.
Operation notes
1. Leave no glass, plastic or water vapor etc. between the unit and target.
2. Keep the product away from the following places to reduce the risk of damaging the device:
A, Environment with vapor and dust;
B, EMF places
(
Electro-magnetic fields: such as arc welding machine, induction heaters
)
;
C, Static environment;
D, Heat shock (caused by abrupt temperature changes, allow 30 minutes for unit to stabilize before use in such
circumstance);
E, Stay away from high temperature objects.
Technical Specifications
Name
Infrared Thermometer
Model
KC-180B-1
Measuring range
-50
℃
~
650
℃
(-58
℉
~
1202
℉
)
Response wavelength
8
~
14μm
Measuring precision
±2
℃
(±3.6
℉
) or ±2%
(
T>0
℃
)
±3
℃
(±5.4
℉
) or ±2% whichever is greater
(
T≤0
℃
)
Repetition
1% of readings or 1
℃
Response time
500mSec
,
95% response
Optical ratio
(
D
:
S
)
12
:
1
Emissivity
0.10
~
1. 00 adjustable
(
default 0.95
)
Display precision
±0.1
℃
Laser wavelength
630
~
660nm
Laser power
<1mW
Laser class
Class II
Laser switch
√
Backlight switch
√
Data hold
√
Maintenance
1. Lens protection: Abrupt temperature changes will cause vapor condensed on the lens, please clean the
lens after the vapor disappears. Blow off loose particles using clean compressed air. Gently brush
remaining debris away with a camel’s hair brush. Carefully wipe the surface with a moist cotton swab.
2. Keep the device clean and tidy, avoid dropping or shocking and keep it away from water. The
housing could be cleaned with wet sponge.