9
1.8 TECHNICAL
DATA
Air-flow rate of pump
Ratiomed AC 20:
18 l/min ± 2 l/min (
low-flow
)
Ratiomed AC 30:
28 l/min ± 2 l/min (
high-flow
)
Vacuum
at least -80 kPa (-800 mbar, 600 mmHg) (
high vacuum
)
conversion factor: 1 kPa ~ 7.5 mmHg
Collection jar
1 l plastic jar (polypropylene)
Suction hose
silicone, transparent, internal diameter 6 mm,
wall thickness 3 mm, length = 1.3 m
Nominal voltage
230 V AC
Mains frequency
50/60 Hz
Power consumption
approx. 80 VA
Current consumption
0.35 A
Dimensions (HxWxD) in mm
290 x 259 + 100 (jar) x 130
Weight (basic unit)
Ratiomed AC 20:
2.9 kg
Ratiomed AC 30:
3.5 kg
Operating time
Ratiomed AC 20:
continuous operation
Ratiomed AC 30:
short-time operation, max. 45 min
Degree of protection acc. to IEC 601-1
Type BF
Risk class acc. to 93/42/EEC, IX
IIa (2a)
Protection class acc. to IEC 601-1
II (completely insulated)
Protection classification acc. to IEC 529
IP 20
CE identification
CE0482
Sound emission
w/o vacuum
with vacuum
Ratiomed AC 20:
approx. 55dB (A)
approx. 56dB (A)
Ratiomed AC 30:
approx. 57dB (A)
approx. 61dB (A)
Ambient conditions
Transport/storage
- 10°C up to + 60°C
Operation
+ 5°C up to + 35°C
5 up to 80% humidity, non-condensing
860 hPa…1060 hPa air pressure
Item numbers
Ratiomed AC 20: REF 8710401
Ratiomed AC 30: REF 8710501