background image

G

 

Safety Warnings

The following safety warnings and precautions 

must be

 read and understood before the 

instrument is used. They must be observed during use.

n

  For safety, only connect the PAT to a supply that is properly earthed. If in doubt, the 

  supply should be checked by a qualified electrician.

n

 

Do not use

 the instrument if there are any signs of damage.

n

  All test leads, probes and clips 

must be

 in good order, clean and with no broken or  

  cracked insulation.

n

  Probes and clips should be held behind the finger guard. 

n

  Test leads not used during a measurement should be disconnected from the appliance  

 tester.

n

  For dual voltage testers, both sockets can be live simultaneously.

n

  Only connect one asset to the PAT during testing.

n

  Tests should be carried out in the order recommended below. An appliance that fails a  

  test should be repaired before further testing is carried out.

Recommended sequence:
1. Visual inspection.
2. Earth bond/ continuity of the protective earth conductor (Class I devices).
3. Insulation test (or earth leakage).
In addition further tests can be performed, but ONLY AFTER 1 & 2 above.
4. Operation test.
5. Leakage test.

n

  Only perform an operational test after the earth bond and insulation tests have been  

  completed, as this test operates at mains voltage.

n

  During testing, ensure no hazard will exist as a result of normal running or under fault  

 conditions.

n

  During testing the unit under test (asset) 

should not

 be touched, other than using the 

  appropriate accessories, as faulty appliances can present a shock hazard.

n

 

Do not

 touch the exposed parts of test leads during tests as hazardous voltages may be  

  present due to potentially faulty appliance.

n

 

Do not

 touch the IEC extension lead socket pins especially during a test, as hazardous  

  voltages may be present due to a potentially faulty appliance.

n

 Assets 

should not

 be routinely flash tested. Where flash testing is required, refer to 

  further guidance on flash testing in the user guide.

n

  Replacement fuses must be of the correct rating and type. Refer to the user guide.

n

  The USB connection marked thus 

   

 should only be used by approved service  

  personnel; nothing should be connected to the USB port during testing.

n

  Only use a NiMH rechargeable 8.4 V PP3 battery, 

do not

 use a non rechargeable type as  

  this could become dangerous if charged by the instrument.

n

  Serviceable fuses 

should

 only be replaced with those that are suitably rated. 

n

  In case of an emergency use an easily accessible power point.

n

  Only use Megger approved accessories with this product.

For service requirements contact:

Megger   

  

 

or 

Megger    

  

 

Archcliffe Road      

 

 

Valley Forge Corporate Centre 

Dover    

 

 

 

2621 Van Buren Avenue 

 

Kent  CT17 9EN   

 

 

Norristown PA 19403 

 

England. 

    U.S.A.

Tel:  +44 (0) 1304 502 243   

 

Tel:  +1 610 676 8579 

 

Fax: +44 (0) 1304 207 342    

 

Fax: +1 610 676 8625

Symbols used on the instrument

F

 

Caution

: risk of electric shock.

G

 

Caution

: refer to accompanying notes. 

c

  Equipment complies with relevant EU Directives.

 

Equipment complies with ‘C tick’ requirements.

 Fused.

 

This equipment should be recycled as electronic waste.

 

HV test lead in unlocked position.

 

HV test lead in locked position.

 

Battery type fitted.

 

DO NOT

 connect to 230 V supply.

Printed in the UK Part No. PAT400_QS_2000-965_en_V03      

   

N13117

Отзывы: