MIT200 series

Digital/Analogue Insulation and continuity testers

MIT200 series

Digital/Analogue Insulation and continuity 

testers

Insulation testing to 1000 MΩ 

Continuity testing at 200 mA down to 0.01 Ω 

Live circuit warning (voltage display) and test 
inhibit 

Digital/analogue display 

Alkaline or rechargeable batteries

–10 °C to +55 °C operating temperature

CATIII 600 V 

Conforms to EN 61557-1

DESCRIPTION

The MIT200 is one of the smallest insulation testers available on 
the market today. With options of a two and three test voltage 
instruments, the MIT200 instruments offer a range of safety and 
operation features. 
The display offers a combination of digital readout and analogue 
display, using Megger’s patented DART display technology, which 
include the benefits of an LCD display, such as robust, clear and 
accurate measurement, with an analogue pointer response for 
evaluating circuit charge and discharge characteristics. 
The instrument housing is in tough ABS, designed to withstand the 
rigours of hard use, and is small enough to drop into your pocket 
when not in use. 
Battery requirements are 6 AA batteries of either standard alkaline 
or nickel metal hydride (NiMH) rechargeable type. A low battery 
warning indicator gives advanced warning of exhausted batteries. 

Continuity testing 

Automatic continuity testing is performed at 200 mA to ensure 
compliance with international requirements. No need to press the 
test button. 
All instruments will measure up to 100 Ω on continuity, of which 
0-10 Ω is performed at greater than 200 mA to meet international 
electrical testing requirements. 
Lead null is possible up with to 9.99 Ω of test lead resistance, 
ensuring the ability to null fused test leads as well as standard leads. 

Continuity buzzer 

A continuity buzzer provides a means of rapid cable testing and 
circuit identification, with voltage protection should you accidentally 
touch a live circuit. 
The buzzer operates at a 2 Ω threshold. 

Insulation testing 

The instruments offer one of two configurations as detailed on page 
2, providing an ideal solution to most low voltage insulation testing 
applications. 
Insulation testing is possible up to 1000 MΩ on all ranges. 
Auto discharge ensures all circuits are safely discharged after testing. 
1000 V insulation test ranges have a high voltage warning prior to 
test voltage being applied. 

Safety 

Every Megger instrument is designed with safety as its primary 
objective. All instruments meet or exceed the requirements of safety 
directive IEC 61010 and EN 61557 for insulation and continuity 
testing 

Default fault meter 

A built-in voltmeter automatically switches in when the instrument is 
connected to a circuit with an a.c. or d.c. voltage greater than 25 V. 

Test inhibit 

Circuits in excess of 25 V will generate a voltage warning. Circuits 
over 50 V will inhibit testing on both continuity and insulation test 
ranges, protecting the operator and the instrument from injury or 
damage. 

600 V CATIII 

The MIT200 series has been designed for use on applications up to 
600 V CAT III.