Functions:
Buzzer ON/OFF
Buzzer threshold:
Buzzer sounds below 2
Ω
. To change limit go to
SETUP
Warnings:
Live circuit test inhibit at 3 V
SETUP options:
Test current:
200 mA
/ 15 mA
Auto reverse testing: On/
Off
Continuity buzzer threshold: 0.5
Ω
, 1
Ω
,
2 Ω
, 5
Ω
, 10
Ω
, 20
Ω
, 50
Ω
and
100
Ω
Insulation measurement [M
Ω
]
MEASUREMENT:
Press TEST to start. Keep pressed until measurement
stabilises.
Release TEST button to discharge circuit.
Test Lock:
Hold down TEST button and press LOCK. To release
press TEST.
Functions:
Insulation “Pass threshold” buzzer ON/OFF
Buzzer threshold:
Sounds above 0.5 M
Ω
, adjustable in SETUP.
Warnings:
Ensure circuit to be tested is disconnected and
isolated.
Live circuit warning at 25 V or 50 V (Default 50 V)
SETUP options:
Insulation buzzer threshold: 0.5 M
Ω
, 1 M
Ω
, 2 M
Ω
,
3 M
Ω
, 4 M
Ω
, 5 M
Ω
, 7 M
Ω
, 10 M
Ω
, 50 M
Ω
, 100 M
Ω
,
500 M
Ω
(Default 0.5 M
Ω
)
RCD testing [ms] –
Measures trip times of RCD
Main
range
knob
(left)
2nd
range
knob
Terminal
connections
Separate
leads
Mains lead
connections
User
action
250 V – 500 V
MFT1710
Not used
Press
and
hold
TEST
button
L-N
L-E
Main
range
knob
(left)
2nd range
knob
Terminal
connections
Separate
leads
Mains lead
connections
User
action
30 mA – 500 mA
MFT1710
See note
See note
See note
Press
TEST
button