10.4.1 Replace batteries and remove isolation tab
The batteries are housed in the lower section of the DLRO2, behind the rotary switch. Looking at the back of the
instrument, the lower (battery) cover incorporates a stand and has the replacement cell types written on it.
NOTE:
If using the DLRO2 for the first time remove and discard the battery isolation tab (3).
1. Remove the test leads and ensure the DLRO2 is switched off.
2. Slacken one captive cross-point screw (1) which is centred at the top of the cover.
3. Battery cover (2) will now lift away from the top.
4. Ease out of lugs at the bottom of the cover.
5. The six batteries (4) can now be lifted out of the compartment.
Caution: Ensure the new cells are replaced in the correct polarity as indicated on the cells and battery
compartment.
Caution: Ensure all batteries are of the same type, do not mix NiMH or Alkaline.
6. Replace all six batteries (4).
7. Replace the battery cover (2) in reverse order to above.
8. Re-secure with screw (1).
9. If the type of batteries has changed (NiMH or Alkaline) ensure the battery technology setting is changed.
Maintenance
www.megger.com
DLRO2
34