For a single test the result is displayed constant in the main
field. No latest results are displayed in the secondary field
unless multiple tests have been made in this sequence.
Note:
The TICK button (
) will clear the previous results in
the secondary fields.
Two previous results appear in the secondary fields, left is the
previous result and right is the result before that.
For auto single tests the continuity indicators are grey,
indicating that to repeat the test the leads must be
disconnected and reconnected again.
If the test has been stopped by the TEST button the continuity
indicators will be grey.
If continuity already is shown on the C and P connectors press
the TEST button to re-start the test.