25
SAFETY
The instruments meet the requirements for double
insulation to IEC 1010-1 (1995), EN 61010-1 (1995) to
Category III*, 300 Volts phase to earth (ground) and 440
Volts phase to phase, without the need for separately
fused test leads. If required, fused test leads are available
as an optional accessory.
* Relates to the transient over-voltages likely to be met in
fixed wiring installations.
Complies with the following parts of EN61557, Electrical
safety in low voltage systems up to 1000V a.c. and 1500V
d.c. – Equipment for testing, measuring or monitoring of
protective measures:-
Part 1 - General requirements
Part 2 - Insulation resistance
Part 4 - Resistance of earth connection and equi-
potential bonding
FUSE
500mA 500V, 32x 6mm Ceramic HBC 10kA minimum.
E.M.C.
The instruments meet 61326-1
POWER SUPPLY
Battery Type:
6x1,5V Alkaline cells
IEC LR6 type or 1.2V NiCd
re-chargeable cells.
Battery Life (typical): 2100 5-sec 1kV insulation tests
3200 5-sec 500V insulation tests
4000 5-sec 250V insulation tests
2700 5-sec continuity tests
4700 5-sec k
Ω
tests
ENVIRONMENTAL CONDITIONS
Operating range:
-5 to +40°C
Operating humidity:
90% RH at 40°C max.
Storage temperature range:
-25 to +65°C
Calibration Temperature:
+20°C
Maximum altitude:
2000m
Dust and water protection:
IP54
Temperature coefficient:
<0,1% per °C
WEIGHT:
742g
DIMENSIONS:
110mm x 220mm x 45mm
CLEANING:
Wipe with a clean cloth
dampened with soapy
water or Isopropyl Alcohol
(IPA)