12
Class I test
(PAT120, 150) using substitute leakage @ 40 V ac
i)
ii)
iii)
Connect BOND lead P1
iv)
Class I @ 500 V
OR
Class I @ 250 V for IT equipment
>2 s
v)
Ensure probe (P1) is connected
vi)
Remove probe (P1)
See note below
vii)
viii) Class I Pass
NOTE:
If the contact symbol
is displayed during the test, the PAT has detected an open circuit load. Ensure the appliance is switched on then press the Class I icon
NOTE
: The PAT100 instruments perform various pre-checks prior to testing to ensure the asset is not short-circuit and is switched on