It is highly recommended to check whether the tight contact obtained or not after 
the first testing of the generator. Unscrew the cable connector; it should be no any 
ozone or burnt smell from the connectors or any burnt traces.  
During  normal  operation,  please  do  not  disconnect  and  connect  again  the  high 
voltage connectors many times to prevent the contacts wearing and contamination. 
Silicon  oil  has  been  already  applied  to  both  connectors  at  the  factory.  It  remains 
inside the mating connector pair during very long time and prevent arcing. 
 

Step 4. 

Attach  semirigid  coaxial  cable  assembly  to  the  output  module  SMA  connector; 
attach  two  20dB  attenuators  and  if  required  SMA-to-BNC  adapter  for  the 
connection  to  oscilloscope  (see  Fig.4).  1:40  coupler  together  with  two  20dB 
attenuators  gives  4000  total  division  coefficient.  10  kV  pulse  divided  by  4000 
times  results  in  2.5V  measured  signal  which  can  be  directly  applied  to  50  Ohm 
impedance oscilloscope input. Some of high bandwidth oscilloscopes allow up to 
1V maximum input pulse amplitude only. Please install one more 20dB attenuator 
(not included) in this case. 
 

 

 
Fig.4.  Output module connection to NRG1001 generator, sample and oscilloscope. 
 
 

 

  

Attach  the  attenuators  just  to  the  oscilloscope  input,  but  not  to  the  SMA 
coupler  output  of  the  module.  This  improve  signal-to-noise  ratio  of  the 
registered signal.  

 

User manual of NRG1001 HV rectangular pulse generator                                  9

 

Содержание NRG1001

Страница 1: ...NANOSECOND HIGH VOLTAGE RECTANGULAR WAVEFORM PULSE GENERATOR NRG1001 USER MANUAL v 0 1 2018 Megaimpulse Ltd...

Страница 2: ...OR ERROR IN THIS MANUAL OR PRODUCT SPECIFICATIONS AND INFORMATION CONTAINED IN THIS MANUAL ARE FURNISHED FOR INFORMATION ONLY AND ARE SUBJECT TO CHANGE AT ANY TIME WITHOUT NOTICE AND SHOULD NOT BE CO...

Страница 3: ...generator 3 Technical specification 5 Front and rear panels view 6 Package content 7 Putting the generator into operation 8 Triggering of the generator 11 SYNC OUT pulse 12 Fuse replacement 13 Warrant...

Страница 4: ...system ensure that the power supply ON OFF toggle is switched off and power supply cable is unplugged before the output cable is connected or disconnected Please connect or disconnect any equipment w...

Страница 5: ...s higher in most of practical applications which gives up to double output pulse voltage on the load There are internal and external triggering operation modes which can be chosen by EXT INT toggle sw...

Страница 6: ...2 General view of NRG1001 nanosecond high voltage rectangular like waveform pulse generator Fig 3 Output modules with integrated 1 40 resistive coupler and with or without 1 3 divider MEGAIMPULSE LTD...

Страница 7: ...1 5 kV 5 0 kV on 50 Ohm matched load Output pulse width 10 ns fixed FWHM Output pulse rise fall times 2 2 ns 4 5 ns 2 5 ns 5 0 ns depending on the pulse amplitude and load impedance Repetition rates f...

Страница 8: ...nnector 3 power supply connector and fuse holder 4 HV ON HV OFF high voltage on off switch 5 EXT INT external internal triggering toggle switch 6 BNC SYNC IN connector 7 AMPLITUDE regulation knob 8 FR...

Страница 9: ...assembly 2 consisting of RG214 U cable special HV coaxial connector output module with integrated 1 3 divider and 1 40 resistive coupler Semirigid coaxial cable assembly with SMA connectors Two 20dB...

Страница 10: ...ou want to operate with depending on the required output pulse amplitude Output module without the divider allows to get 3kV 10kV pulse amplitude on high impedance load output module with 1 3 divider...

Страница 11: ...module SMA connector attach two 20dB attenuators and if required SMA to BNC adapter for the connection to oscilloscope see Fig 4 1 40 coupler together with two 20dB attenuators gives 4000 total divis...

Страница 12: ...m frequency about 1 Hz Connect power supply cable to wall power outlet Switch on the generator by power switch on the rear panel The internal fan should start to rotate Step 7 Switch on the high volta...

Страница 13: ...mp in HV ON HV OFF switch lights on The maximum allowable pulse repetition rate in this mode is limited by FREQUENCY regulation knob at the front panel of the generator Therefore internal generator fr...

Страница 14: ...h impedance load The SYNC OUT pulse precedes output HV pulse by about 300 ns The oscillograms of SYNC OUT pulse Channel 1 yellow line and output HV pulse Channel 4 magenta line are shown in Fig 5 Fig...

Страница 15: ...nnector Please use flat screwdriver or other suitable tool to remove the fuse holder see Fig 6 Fig 6 Removing of the fuse holder by flat screwdriver There are two fuses in the fuse holder including on...

Страница 16: ...ent to determine the warranty period and condition The generator has warranty seals on the front and rear panels see Fig 8 Removing of the seals terminates the warranty Fig 8 Warranty seals on the fro...

Отзывы: