Care and Cleaning 

The probe tip (lens) is the most fragile part of the thermometer. It is essential you keep it clean at all times and avoid exposure 
to hot steam, water, and dirt. 

Cleaning the lens: 

Very gently wipe the surface with a cotton swab or soft cloth moistened with an alcohol solution. And allow to dry out com-
pletely. Please avoid cleaning the lens unless strictly necessary. 

Cleaning the body: 

Use a soft, dry cloth to clean the thermometer display and body. If very dirty, use a soft cloth with an alcohol solution. 

DO NOT USE ABRASIVE CLEANERS, SOLUTIONS, OR SUBMERGE IN WATER. 

 

Trouble-shooting

 

 

Fault 

Checklist 

Solution 

No response / auto-
matically reset 

1. Check the batteries 

2. Poor battery contact 

1. Replace the batteries 

2. Check the battery polarity is cor-
rect 

The thermometer is 
showing the symbol 
“Hi” 

In forehead measurement mode “Hi” > 43.0°C; 

1. The temperature readings are too close together 

2. The device is measuring interference i.e. environment 
temperature, sunlight, fireplace heat. 

In surface measurement mode “Hi” > 100.0°C; 

1. The temperature readings are too close together 

2. The object temperature is higher than 100.0°. 

Please leave the status for at least 
30 minutes then re-measure after 
adjusting as necessary. 

The thermometer is 
showing the symbol 
“Lo” 

In forehead measurement mode “Lo” < 32.0°C; 

1. The measuring distance is too far 

2. The device is measuring interference i.e. environment 
temperature, air conditioner air. 

In surface measurement mode “Lo” < 0°C; 

1. The measuring distance is too far 

2. There is water vapour condensation on the lens 

Please leave the status for at least 
30 minutes then re-measure after 
adjusting as necessary. 

“Err” Error symbol is 
displaying 

1. The ambient temperature is beyond the range of 
measurement (10.0°C—40.0°C) 

2. The sensor or hardware is damaged 

1. Keep the thermometer where it is 
and allow 30 minutes to adjust to 
the ambient temperature or change 
locations 

2. Contact manufacturer to trouble-
shoot 

Low battery symbol 

The battery is low 

Replace the batteries 