Section 5: Service Instructions – Gas Spring
Replacement MedViron 46” Overbed Table
This manual is intended for trained professionals for adjustment of the MedViron Overbed Table. Any
action not outlined in this manual may result in voiding of the warranty.
Before moving to this step, first attempt the adjustment in Section 1.
The overbed table contains a single locking gas spring for height adjustment. This gas spring keeps
the table from being lowered until the handle is activated, but allows free rise. Gas springs can lose
force over time, which will cause the table to descend faster when activated.
Tools required
#2 & #3 Phillips Screwdriver (Powered driver not recommended except where noted)
7/16” socket or wrench
Block (provided)
5.0
Replacing the gas spring
5.1
Lift table to the highest adjustment position. Remove shelf or drawer pedestal following
instructions in Section 6.